INSTRUCTIONS EVIS EXERA II GASTROINTESTINAL VIDEOSCOPE OLYMPUS GIF TYPE N180 OLYMPUS GIF TYPE Q180 OLYMPUS GIF TYPE H180 EVIS EXERA II COLONOVIDEOSCOPE OLYMPUS CF TYPE Q180AL/I OLYMPUS CF TYPE H180AL/I OLYMPUS PCF TYPE Q180AL/I Refer to the endoscope’s companion manual, the “REPROCESSING MANUAL” whose cover lists the model of your endoscope, for reprocessing information.
Inspection and connection of ancillary equipment ......Inspection of the endoscopic system ..........Chapter 4 Operation ..............Insertion ..................Using endo-therapy accessories............. Withdrawal of the endoscope............Transportation of the endoscope ............ EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
Page 4
Contents Chapter 5 Troubleshooting ............Troubleshooting guide ..............Withdrawal of the endoscope with an abnormality......Returning the endoscope for repair..........Appendix..................System chart .................... EMC information..................EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
Refer to instructions. Endoscope TYPE BF applied part Single use only Lot number Manufacturer Authorized representative in the European Community EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
Important Information — Please Read Before Use Intended use These instruments have been designed to be used with an Olympus video system center, light source, documentation equipment, video monitor, endo-therapy accessories such as a biopsy forceps and other ancillary equipment.
It also accompanied the endoscope at shipment. Keep this and all related instruction manuals in a safe, accessible location. If you have any questions or comments about any information in this manual, please contact Olympus. EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
After using this instrument, reprocess and store it according to the instructions given in the endoscope’s companion reprocessing manual. Improper and/or incomplete reprocessing or storage can present an infection-control risk, cause equipment damage or reduce performance. EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
Equipment which has been disassembled, repaired, altered, changed or modified by persons other than Olympus’ own authorized service personnel is excluded from Olympus’ limited warranty and is not warranted by Olympus in any manner. Signal words...
Before proceeding, always be sure to confirm that transnasal insertion is possible with the patient. Otherwise, operator and/or patient injury can result, or the endoscope could become lodged and be difficult to withdraw. EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
Page 11
Never perform angulation control forcibly or abruptly. Never forcefully pull, twist or rotate the angulated bending section. Patient injury, bleeding and/or perforation can result. It may also become impossible to straighten the bending section during an examination. EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
Page 12
Before use, be sure to read the instruction manual for the splinting tube to fully understand its characteristics. EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
Page 13
CCD. • The endoscope’s remote switches cannot be removed from the control section. Pressing, pulling or twisting them with excessive force can break the switches and/or may cause water leaks. EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
Page 14
This endoscope contains a memory chip that stores information about the endoscope and communicates this information to the video system center CV-160 and CV-180. EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
• Inserting or withdrawing the endoscope, feeding air, applying suction or operating the bending section without a clear endoscopic image may cause patient injury, bleeding and/or perforation. EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
Match all items in the package with the components shown below. Inspect each item for damage. If the instrument is damaged, a component is missing or you have any questions, do not use the instrument; immediately contact Olympus. This instrument was not disinfected or sterilized before shipment.
Page 17
Chapter 1 Checking the Package Contents GIF-N180 Endoscope GIF-Q180, GIF-H180 Endoscope CF-Q180AL/I, CF-H180AL/I, PCF-Q180AL/I Endoscope EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
Page 18
(MAJ-855 for endoscopes with (MA-474, MB-142 for (MB-142 for GIF-Q180, auxiliary water feeding only) GIF-N180) (1 pc each) GIF-H180) (2 pcs) Instructions Reprocessing manual Operation manual (leaflet type, for endoscopes with flexibility adjustment only) EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
Page 19
Auxiliary water tube GIF-H180) (2 pcs) GIF-N180) (1 pc each) (MAJ-855 for endoscopes with auxiliary water feeding only) Operation manual Reprocessing manual Instructions (leaflet type, for endoscopes with flexibility adjustment only) EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
1. Suction connector 5. Electrical connector 15. Chain connector Air pipe 3. Air supply connector 3. Water supply connector Light guide Product name and serial number Contact pins 4. Endoscope connector EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
Page 21
14. Color code Biopsy valve (MB-358) Boot 10. Instrument channel Instrument channel port 11. Insertion tube limit mark Working length Distal end 13. Remote switches 1 to 4 12. Bending section Insertion tube EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
Page 22
1. Suction connector 5. Electrical connector 2. S-cord connector mount Air pipe 3. Air supply connector 3. Water supply connector Light guide Product name and serial number Contact pins 4. Endoscope connector EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
Page 23
14. Color code Biopsy valve (MB-358) Boot 10. Instrument channel Instrument channel port 11. Insertion tube limit mark Working length 13. Remote switches 1 to 4 Distal end 12. Bending section Insertion tube EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
Page 24
Auxiliary water inlet cap (MAJ-215) Air pipe 2. S-cord connector mount 3. Air supply connector 3. Water supply connector Light guide Product name and serial number Contact pins 4. Endoscope connector EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
Page 25
10. Instrument channel 18. Flexibility adjustment ring Instrument channel port Marks Boot 11. Insertion tube limit mark Working length Insertion tube 13. Remote switches 1 to 4 Distal end 12. Bending section EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
Connects the endoscope to the suction tube of the suction pump. 2. S-cord connector mount (except GIF-N180) Connects the endoscope with the Olympus electrosurgical unit via the S-cord. The S-cord conducts leakage current from the endoscope to the electrosurgical unit. To connect the S-cord, refer to the instruction manual for the electrosurgical unit.
Page 27
“ ” mark (“3” corresponds to the most-rigid condition). As the ring is turned from “ ” to “3”, the insertion tube’s flexibility gradually decreases. EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
Transportation and Ambient temperature –47 to 70°C (–52.6 to 158°F) storage Relative humidity 10 – 95% environment Atmospheric pressure 700 – 1060 hPa (0.7 – 1.1 kgf/cm (10.2 – 15.4 psia) EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
Page 29
Available ∗1 GIF-N180 cannot be used to perform high-frequency cauterization or laser cauterization. ∗2 NBI stands for Narrow Band Imaging. For more details, refer to the instruction manual of the CV-180. EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
Page 30
UP 210°, DOWN 90°, section RIGHT 100°, LEFT 100° Total length 1345 mm ∗1 Available ∗1 NBI stands for Narrow Band Imaging. For more details, refer to the instruction manual of the CV-180. EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
Page 31
UP 210°, DOWN 90°, section RIGHT 100°, LEFT 100° Total length 1345 mm ∗1 Available ∗1 NBI stands for Narrow Band Imaging. For more details, refer to the instruction manual of the CV-180. EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
Page 32
RIGHT 160°, LEFT 160° Total length L: 2005 mm I: 1655 mm ∗1 Available ∗1 NBI stands for Narrow Band Imaging. For more details, refer to the instruction manual of the CV-180. EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
Page 33
RIGHT 160°, LEFT 160° Total length L: 2005 mm I: 1655 mm ∗1 Available ∗1 NBI stands for Narrow Band Imaging. For more details, refer to the instruction manual of the CV-180. EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
Page 34
RIGHT 160°, LEFT 160° Total length L: 2005 mm I: 1655 mm ∗1 Available ∗1 NBI stands for Narrow Band Imaging. For more details, refer to the instruction manual of the CV-180. EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
Page 35
1. Year of 2512345 manufacture The last digit of the year of manufacture is the second digit of the serial number. Degree of TYPE BF applied part protection against electric shock EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
ETO gas sterilized; doing so may damage them. If the water-resistant cap is connected to the endoscope by the chain, be sure to remove the chain and the water-resistant cap from the endoscope before ETO gas sterilization. EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
Page 37
Confirm that the connecting plate is securely attached to the foot of the venting connector and can be smoothly rotated. Place the hole on the fitting over the endoscope’s S-cord connector mount or the chain connector of the GIF-N180 (see Figure 2.3). EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
Page 38
Figure 2.3 The instructions on the remaining pages of this manual are given under the assumption that the chain for the water-resistant cap is detached from the endoscope. EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
If any irregularities are suspected after inspection, follow the instructions as described in Chapter 5, “Troubleshooting”. If this instrument malfunctions, do not use it. Return it to Olympus for repair as described in Section 5.3, “Returning the endoscope for repair” on page 86.
Mouthpiece (for GIF models only) Endo-therapy accessories Auxiliary water tube (for endoscopes with auxiliary water feeding only) • Paper towels • Trays • Lint-free cloths • Personal protective equipment Figure 3.1 EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
When inspecting endoscopes with flexibility adjustment, perform the test with the insertion tube at both its most-flexible and most-rigid settings (for endoscopes with flexibility adjustment only). EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
Page 42
Confirm that the marks (“ ”, “1”, “2”, “3”) on the flexibility adjustment ring and the “ ” mark at the bottom of the grip section are clearly visible (see Figure 3.4). Figure 3.4 EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
Page 43
Figure 3.5. Confirm that the actual flexibility changes according to the flexibility adjustment setting. 50 cm 30 cm Figure 3.5 EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
Page 44
(the GIF-N180 has only the UP/DOWN angulation control knob). Confirm that the bending section angulates smoothly and correctly, that maximum angulation can be achieved, and that the bending section returns to its neutral position. EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
Page 45
When the UP/DOWN angulation control knob is turned to its neutral position as shown in Figure 3.7, confirm that the bending section returns smoothly to an approximately straight condition (for GIF-N180 only). Figure 3.7 EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
Page 46
RIGHT/LEFT angulation control knob is released. Confirm that the bending section straightens out when the RIGHT/LEFT angulation lock is turned in the “F ” direction and the RIGHT/LEFT angulation control knob is released. EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
Confirm that the valves are not deformed or cracked (see Figures 3.8 and 3.9). Check for excessive scratching or tears in the air/water valve’s seals (see Figure 3.8). Hole Spring Seals Skirt Hole Air/water valve (MH-438) Figure 3.8 EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
Page 48
Confirm that the slit and hole on the biopsy valve have no splits, cracks, deformation, discoloration or other damage (see Figure 3.10). Normal Abnormal Slit Discoloration Hole Splits, Cracks Main body Figure 3.10 EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
Page 49
If irregularities are observed, replace it with a new one as described in “Attaching the auxiliary water inlet cap (for endoscopes with auxiliary water feeding only)” on page 49. Auxiliary water inlet Auxiliary water inlet cap Endoscope connector Figure 3.12 EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
Page 50
Confirm that the mouthpiece is free from cracks, deformation or discoloration (see Figure 3.14). Using your fingers, check for excessive scratching or other irregularities on all surfaces of the mouthpiece (see Figure 3.14). EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
Attach the suction valve to the suction cylinder of the endoscope (see Figures 3.15 and 3.16). Confirm that the valve fits properly without any bulging of the skirt. Also confirm that the valve cannot be rotated. EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
Page 52
Confirm that the valve fits properly without any bulging of the skirt. Air/water valve Suction valve Skirt Suction cylinder Air/water cylinder Figure 3.16 The air/water valve may stick at first, but it should operate smoothly after it is depressed a few times. EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
Page 53
If the auxiliary water inlet cap is not attached, attach the fitting ring to the auxiliary water inlet on the endoscope connector (see Figure 3.18). Fitting ring Auxiliary water inlet cap Auxiliary water inlet Endoscope connector Figure 3.18 EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
Connect the water container’s connection adapter to the air supply connector and water supply connector (see Figure 3.19). Confirm that the water container’s connection adapter fits properly and that it cannot be rotated. EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
Page 55
Turn the connector of the videoscope cable clockwise until it stops (see Figure 3.20). Confirm that the mark on the videoscope cable is aligned with mark 2 on the endoscope connector. EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
Page 56
(for endoscopes with auxiliary water feeding only, see Figure 3.22). Auxiliary water tube Auxiliary water inlet Auxiliary water inlet cap Endoscope connector Figure 3.22 EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
10 cm and confirm that no air bubbles are emitted when the air/water valve is not operated. Cover the hole in the air/water valve with your finger and confirm that air bubbles are continuously emitted from the air/water nozzle. EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
Page 58
Observe the endoscopic image and confirm that water flows on the entire objective lens. Release the air/water valve. Observe the endoscopic image and confirm that the emission of water stops and that the valve returns smoothly to its original position. EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
Page 59
Remove the distal end of the endoscope from the water. Depress the suction valve and aspirate air for a few seconds to remove any water from the instrument channel. EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
Page 60
Disconnect the water tube from the water pump or the syringe from the luer port of the auxiliary water tube. Make sure that no water leaks from the luer port of the auxiliary water tube and/or the distal end of the insertion tube. EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
Page 61
This could cause damage to and/or malfunction of the equipment. Syringe or water tube from a water pump Luer port Auxiliary water tube Connecting end Figure 3.23 EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
In this case, the distal end of the endoscope can become hot and could cause operator and/or patient burns (when using a light source other than CLV-180). EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
Page 63
If any of the following phenomena occur during an examination, immediately stop the examination and withdraw the endoscope from the patient as described in Section 5.2, “Withdrawal of the endoscope with an abnormality” on page 83. EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
Page 64
(for endoscopes with flexibility adjustment only). EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
Page 65
If this vapor continues to interfere with the examination, remove the endoscope, wipe the distal end with a lint-free cloth moistened with 70% ethyl or isopropyl alcohol, reinsert the endoscope and continue the examination. EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
If resistance to insertion is felt, or the patient reports pain, stop insertion immediately. Otherwise, operator and/or patient injury can result or the endoscope could become lodged and be difficult to withdraw. EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
Page 67
10 cm or less from the junction of the boot. Insertion tube damage can occur (see Figures 4.2 and 4.3). > 10 cm Figure 4.2 Endoscopes with flexibility adjustment > 10 cm Figure 4.3 EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
Page 68
Operate the angulation control knobs as necessary to guide the distal end for insertion and observation. The endoscope’s angulation locks are used to hold the angulated distal end in position. EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
Page 69
In this case, straighten the insertion tube as much as possible before attempting to increase the rigidity. Failure to do so may cause patient pain, injury, bleeding and/or perforation. EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
Page 70
EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
Page 71
Cover the air/water valve’s hole to feed air from the air/water nozzle at the distal end (see Figure 4.5). Depress the air/water valve to feed water onto the objective lens (see Figure 4.5). Suction valve Air/water valve Figure 4.5 EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
Page 72
Depress the suction valve to aspirate excess fluid or other debris obscuring the endoscopic image (see Figure 4.5). Performing both air feeding and suction at the same time sometimes makes it easier to remove water droplets from the objective lens surface. EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
Page 73
∗1 NBI stands for Narrow Band Imaging. For more details, refer to the instruction manual of the CV-180. EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
If the distal end of an endo-therapy accessory is not visible in the endoscopic image, do not open the distal end or extend the needle of the instrument. This could cause patient injury, bleeding, perforation and/or equipment damage. EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
Page 75
When not using an endo-therapy accessory, attach the cap to the main body of the biopsy valve. EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
Page 76
Hold the endo-therapy accessory approximately 4 cm from the biopsy valve and advance it slowly and straight into the biopsy valve using short strokes while observing the endoscopic image. EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
Page 77
Take care not to cause tissue trauma. Withdraw the endo-therapy accessory slowly while the tip of the endo-therapy accessory is closed and/or retracted into its sheath. EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
Page 78
Prepare, inspect and connect the electrosurgical unit and electrosurgical accessories as described in their instruction manuals. The application of high-frequency current may interfere with the endoscopic image. This does not indicate a malfunction. EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
Page 79
Do not use a damaged laser probe. A laser probe with a damaged sheath or distal end may cause patient injury and/or equipment damage. Prepare, inspect and connect the laser unit and laser probe as described in their instruction manuals. EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
(for CF/PCF models only). Carefully withdraw the endoscope while observing the endoscopic image. Remove the mouthpiece from the patient’s mouth (for GIF models only). EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
(see Figure 4.7). Figure 4.7 EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
Page 82
Putting the endoscope in the carrying case while the insertion tube is rigid could damage the endoscope (for endoscopes with flexibility adjustment only). EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
Some problems that appear to be malfunctions may be correctable by referring to Section 5.1, “Troubleshooting guide”. If the problem cannot be resolved by the described remedial action, stop using the endoscope and send it to Olympus for repair. Olympus does not repair accessory parts. If an accessory part becomes damaged, contact Olympus to purchase a replacement.
Page 84
Replace it with a new one. damaged. The air/water An incorrect air/water Use a correct air/water valve. valve cannot be valve is used. attached. The air/water valve is Replace it with a new one. damaged. EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
Page 85
An image is An incompatible video Select a compatible video system center. abnormal. system center is being used. An incompatible light Select a compatible light source. source is being used. EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
Page 86
Operate the correct remote switch. switch does not switch is operated. work. The remote switch Set the remote switch function correctly as function has been set described in the video system center’s improperly. instruction manual. EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
(for CF/PCF models only). Carefully withdraw the endoscope while observing the endoscopic image. Remove the mouthpiece from the patient’s mouth (for GIF models only). EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
Page 88
(for CF/PCF models only). Carefully withdraw the endoscope while observing the endoscopic image. Remove the mouthpiece from the patient’s mouth (for GIF models only). EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
Page 89
(for CF/PCF models only). Release the angulation control knobs and carefully withdraw the endoscope. Remove the mouthpiece from the patient’s mouth (for GIF models only). EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
Olympus. Before returning the endoscope for repair, contact Olympus. With the endoscope, include a description of the malfunction or damage and the name and telephone number of the individual at your location who is most familiar with the problem.
For further details, contact Olympus. If combinations of equipment other than those shown below are used, the full responsibility is assumed by the medical treatment facility. EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
Page 92
EVIS video system center ∗ ∗ center (CV-160 EVIS EXERA II video system (CV-140 center (CV-180) EVIS universal light source EVIS EXERA light source EVIS EXERA II light source (CLV-U40) (CLV-160) (CLV-180) EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
Page 93
(for CF/PCF models only) Sclerotherapy tube (for GIF models only) ∗ Not compatible with GIF-N180 Sclerotherapy balloon ∗ Use a non-flammable gas. (for GIF models except GIF-N180) ∗ Not compatible with GIF-H180 and CF-H180AL/I EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
Page 94
Water-resistant (BW-201T) (MAJ-1339) cap (MH-553) Channel-opening Injection tube Suction cleaning cleaning brush (MH-946) adapter (MH-856) (MH-507) Chain for water-resistant cap (MAJ-1119) ∗ These products may not be available in some areas. EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
Page 95
Accessories (for CF/PCF models only) Splinting tube ∗1 ST-C3 ST-C3S ST-C5 Endoscope ST-C8 CF-Q180AL/I – CF-H180AL/I – PCF-Q180AL/I – – – applicable – not applicable ∗1 This accessory may not be available in some areas. EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
Page 100
Standard type Spray type Endoscope GIF-N180 PW-2L-1 PW-6P-1 GIF-Q180 PW-1L-1 PW-5L-1 GIF-H180 PW-1L-1 PW-5L-1 CF-Q180AL PW-1V-1 PW-5V-1 CF-Q180AI PW-1V-1 PW-5V-1 CF-H180AL PW-1V-1 PW-5V-1 CF-H180AI PW-1V-1 PW-5V-1 PCF-Q180AL PW-1V-1 PW-5V-1 PCF-Q180AI PW-1V-1 PW-5V-1 EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
Page 101
NM-200U CD-11Z, CD-110U CF-H180AI HX-400U-30 NM-4U-1 NM-200U CD-11Z, CD-110U PCF-Q180AL HX-400U-30 NM-4U-1 NM-200U CD-21Z, CD-120U PCF-Q180AI HX-400U-30 NM-4U-1 NM-200U CD-21Z, CD-120U ∗1 These accessories may not be available in some areas. EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
Page 102
D-201-14304 CF-Q180AI – – D-201-14304 CF-H180AL – – D-201-15004 CF-H180AI – – D-201-15004 PCF-Q180AL – MAJ-297 D-201-12704 PCF-Q180AI – MAJ-297 D-201-12704 ∗1 These accessories may not be available in some areas. EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
Voltage Complies This instrument stabilizes its own radio variability and has no affect fluctuations/flicker such as flicker in lighting apparatus. emissions IEC 61000-3-3 EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
Page 106
It is recommended to use this instrument by (50/60 Hz) maintaining enough distance from any magnetic field equipment that operates with high current. IEC 61000-4-8 is the AC mains power supply prior to application of the test level. EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
Page 107
• Electromagnetic interference may occur on this instrument near a high-frequency electrosurgical equipment and/or other equipment marked with the following symbol: EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
Page 108
EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...