EVIS EXERA II GASTROINTESTINAL VIDEOSCOPE
Refer to the endoscope's companion manual, the "REPROCESSING MANUAL" whose
cover lists the model of your endoscope, for reprocessing information.
USA: CAUTION: Federal law restricts this device to sale by or on the order of a
physician.
INSTRUCTIONS
OLYMPUS GIF TYPE N180
OLYMPUS GIF TYPE Q180
OLYMPUS GIF TYPE H180
EVIS EXERA II COLONOVIDEOSCOPE
OLYMPUS CF TYPE Q180AL/I
OLYMPUS CF TYPE H180AL/I
OLYMPUS PCF TYPE Q180AL/I

Advertisement

Table of Contents
loading

Summary of Contents for Olympus Evis Exera II

  • Page 1 INSTRUCTIONS EVIS EXERA II GASTROINTESTINAL VIDEOSCOPE OLYMPUS GIF TYPE N180 OLYMPUS GIF TYPE Q180 OLYMPUS GIF TYPE H180 EVIS EXERA II COLONOVIDEOSCOPE OLYMPUS CF TYPE Q180AL/I OLYMPUS CF TYPE H180AL/I OLYMPUS PCF TYPE Q180AL/I Refer to the endoscope’s companion manual, the “REPROCESSING MANUAL” whose cover lists the model of your endoscope, for reprocessing information.
  • Page 3: Table Of Contents

    Inspection and connection of ancillary equipment ......Inspection of the endoscopic system ..........Chapter 4 Operation ..............Insertion ..................Using endo-therapy accessories............. Withdrawal of the endoscope............Transportation of the endoscope ............ EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
  • Page 4 Contents Chapter 5 Troubleshooting ............Troubleshooting guide ..............Withdrawal of the endoscope with an abnormality......Returning the endoscope for repair..........Appendix..................System chart .................... EMC information..................EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
  • Page 5: Symbols

    Refer to instructions. Endoscope TYPE BF applied part Single use only Lot number Manufacturer Authorized representative in the European Community EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
  • Page 6: Important Information - Please Read Before Use

    Important Information — Please Read Before Use Intended use These instruments have been designed to be used with an Olympus video system center, light source, documentation equipment, video monitor, endo-therapy accessories such as a biopsy forceps and other ancillary equipment.
  • Page 7: Applicability Of Endoscopy And Endoscopic Treatment

    It also accompanied the endoscope at shipment. Keep this and all related instruction manuals in a safe, accessible location. If you have any questions or comments about any information in this manual, please contact Olympus. EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
  • Page 8: User Qualifications

    After using this instrument, reprocess and store it according to the instructions given in the endoscope’s companion reprocessing manual. Improper and/or incomplete reprocessing or storage can present an infection-control risk, cause equipment damage or reduce performance. EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
  • Page 9: Spare Equipment

    Equipment which has been disassembled, repaired, altered, changed or modified by persons other than Olympus’ own authorized service personnel is excluded from Olympus’ limited warranty and is not warranted by Olympus in any manner. Signal words...
  • Page 10: Warnings And Cautions

    Before proceeding, always be sure to confirm that transnasal insertion is possible with the patient. Otherwise, operator and/or patient injury can result, or the endoscope could become lodged and be difficult to withdraw. EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
  • Page 11 Never perform angulation control forcibly or abruptly. Never forcefully pull, twist or rotate the angulated bending section. Patient injury, bleeding and/or perforation can result. It may also become impossible to straighten the bending section during an examination. EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
  • Page 12 Before use, be sure to read the instruction manual for the splinting tube to fully understand its characteristics. EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
  • Page 13 CCD. • The endoscope’s remote switches cannot be removed from the control section. Pressing, pulling or twisting them with excessive force can break the switches and/or may cause water leaks. EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
  • Page 14 This endoscope contains a memory chip that stores information about the endoscope and communicates this information to the video system center CV-160 and CV-180. EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
  • Page 15: Examples Of Inappropriate Handling

    • Inserting or withdrawing the endoscope, feeding air, applying suction or operating the bending section without a clear endoscopic image may cause patient injury, bleeding and/or perforation. EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
  • Page 16: Chapter 1 Checking The Package Contents

    Match all items in the package with the components shown below. Inspect each item for damage. If the instrument is damaged, a component is missing or you have any questions, do not use the instrument; immediately contact Olympus. This instrument was not disinfected or sterilized before shipment.
  • Page 17 Chapter 1 Checking the Package Contents GIF-N180 Endoscope GIF-Q180, GIF-H180 Endoscope CF-Q180AL/I, CF-H180AL/I, PCF-Q180AL/I Endoscope EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
  • Page 18 (MAJ-855 for endoscopes with (MA-474, MB-142 for (MB-142 for GIF-Q180, auxiliary water feeding only) GIF-N180) (1 pc each) GIF-H180) (2 pcs) Instructions Reprocessing manual Operation manual (leaflet type, for endoscopes with flexibility adjustment only) EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
  • Page 19 Auxiliary water tube GIF-H180) (2 pcs) GIF-N180) (1 pc each) (MAJ-855 for endoscopes with auxiliary water feeding only) Operation manual Reprocessing manual Instructions (leaflet type, for endoscopes with flexibility adjustment only) EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
  • Page 20: Chapter 2 Instrument Nomenclature And Specifications

    1. Suction connector 5. Electrical connector 15. Chain connector Air pipe 3. Air supply connector 3. Water supply connector Light guide Product name and serial number Contact pins 4. Endoscope connector EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
  • Page 21 14. Color code Biopsy valve (MB-358) Boot 10. Instrument channel Instrument channel port 11. Insertion tube limit mark Working length Distal end 13. Remote switches 1 to 4 12. Bending section Insertion tube EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
  • Page 22 1. Suction connector 5. Electrical connector 2. S-cord connector mount Air pipe 3. Air supply connector 3. Water supply connector Light guide Product name and serial number Contact pins 4. Endoscope connector EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
  • Page 23 14. Color code Biopsy valve (MB-358) Boot 10. Instrument channel Instrument channel port 11. Insertion tube limit mark Working length 13. Remote switches 1 to 4 Distal end 12. Bending section Insertion tube EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
  • Page 24 Auxiliary water inlet cap (MAJ-215) Air pipe 2. S-cord connector mount 3. Air supply connector 3. Water supply connector Light guide Product name and serial number Contact pins 4. Endoscope connector EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
  • Page 25 10. Instrument channel 18. Flexibility adjustment ring Instrument channel port Marks Boot 11. Insertion tube limit mark Working length Insertion tube 13. Remote switches 1 to 4 Distal end 12. Bending section EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
  • Page 26: Endoscope Functions

    Connects the endoscope to the suction tube of the suction pump. 2. S-cord connector mount (except GIF-N180) Connects the endoscope with the Olympus electrosurgical unit via the S-cord. The S-cord conducts leakage current from the endoscope to the electrosurgical unit. To connect the S-cord, refer to the instruction manual for the electrosurgical unit.
  • Page 27 “ ” mark (“3” corresponds to the most-rigid condition). As the ring is turned from “ ” to “3”, the insertion tube’s flexibility gradually decreases. EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
  • Page 28: Specifications

    Transportation and Ambient temperature –47 to 70°C (–52.6 to 158°F) storage Relative humidity 10 – 95% environment Atmospheric pressure 700 – 1060 hPa (0.7 – 1.1 kgf/cm (10.2 – 15.4 psia) EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
  • Page 29 Available ∗1 GIF-N180 cannot be used to perform high-frequency cauterization or laser cauterization. ∗2 NBI stands for Narrow Band Imaging. For more details, refer to the instruction manual of the CV-180. EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
  • Page 30 UP 210°, DOWN 90°, section RIGHT 100°, LEFT 100° Total length 1345 mm ∗1 Available ∗1 NBI stands for Narrow Band Imaging. For more details, refer to the instruction manual of the CV-180. EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
  • Page 31 UP 210°, DOWN 90°, section RIGHT 100°, LEFT 100° Total length 1345 mm ∗1 Available ∗1 NBI stands for Narrow Band Imaging. For more details, refer to the instruction manual of the CV-180. EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
  • Page 32 RIGHT 160°, LEFT 160° Total length L: 2005 mm I: 1655 mm ∗1 Available ∗1 NBI stands for Narrow Band Imaging. For more details, refer to the instruction manual of the CV-180. EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
  • Page 33 RIGHT 160°, LEFT 160° Total length L: 2005 mm I: 1655 mm ∗1 Available ∗1 NBI stands for Narrow Band Imaging. For more details, refer to the instruction manual of the CV-180. EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
  • Page 34 RIGHT 160°, LEFT 160° Total length L: 2005 mm I: 1655 mm ∗1 Available ∗1 NBI stands for Narrow Band Imaging. For more details, refer to the instruction manual of the CV-180. EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
  • Page 35 1. Year of 2512345 manufacture The last digit of the year of manufacture is the second digit of the serial number. Degree of TYPE BF applied part protection against electric shock EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
  • Page 36: Attaching The Chain For Water-Resistant Cap (Maj-1119)

    ETO gas sterilized; doing so may damage them. If the water-resistant cap is connected to the endoscope by the chain, be sure to remove the chain and the water-resistant cap from the endoscope before ETO gas sterilization. EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
  • Page 37 Confirm that the connecting plate is securely attached to the foot of the venting connector and can be smoothly rotated. Place the hole on the fitting over the endoscope’s S-cord connector mount or the chain connector of the GIF-N180 (see Figure 2.3). EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
  • Page 38 Figure 2.3 The instructions on the remaining pages of this manual are given under the assumption that the chain for the water-resistant cap is detached from the endoscope. EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
  • Page 39: Chapter 3 Preparation And Inspection

    If any irregularities are suspected after inspection, follow the instructions as described in Chapter 5, “Troubleshooting”. If this instrument malfunctions, do not use it. Return it to Olympus for repair as described in Section 5.3, “Returning the endoscope for repair” on page 86.
  • Page 40: Preparation Of The Equipment

    Mouthpiece (for GIF models only) Endo-therapy accessories Auxiliary water tube (for endoscopes with auxiliary water feeding only) • Paper towels • Trays • Lint-free cloths • Personal protective equipment Figure 3.1 EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
  • Page 41: Inspection Of The Endoscope

    When inspecting endoscopes with flexibility adjustment, perform the test with the insertion tube at both its most-flexible and most-rigid settings (for endoscopes with flexibility adjustment only). EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
  • Page 42 Confirm that the marks (“ ”, “1”, “2”, “3”) on the flexibility adjustment ring and the “ ” mark at the bottom of the grip section are clearly visible (see Figure 3.4). Figure 3.4 EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
  • Page 43 Figure 3.5. Confirm that the actual flexibility changes according to the flexibility adjustment setting. 50 cm 30 cm Figure 3.5 EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
  • Page 44 (the GIF-N180 has only the UP/DOWN angulation control knob). Confirm that the bending section angulates smoothly and correctly, that maximum angulation can be achieved, and that the bending section returns to its neutral position. EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
  • Page 45 When the UP/DOWN angulation control knob is turned to its neutral position as shown in Figure 3.7, confirm that the bending section returns smoothly to an approximately straight condition (for GIF-N180 only). Figure 3.7 EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
  • Page 46 RIGHT/LEFT angulation control knob is released. Confirm that the bending section straightens out when the RIGHT/LEFT angulation lock is turned in the “F ” direction and the RIGHT/LEFT angulation control knob is released. EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
  • Page 47: Preparation And Inspection Of Accessories

    Confirm that the valves are not deformed or cracked (see Figures 3.8 and 3.9). Check for excessive scratching or tears in the air/water valve’s seals (see Figure 3.8). Hole Spring Seals Skirt Hole Air/water valve (MH-438) Figure 3.8 EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
  • Page 48 Confirm that the slit and hole on the biopsy valve have no splits, cracks, deformation, discoloration or other damage (see Figure 3.10). Normal Abnormal Slit Discoloration Hole Splits, Cracks Main body Figure 3.10 EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
  • Page 49 If irregularities are observed, replace it with a new one as described in “Attaching the auxiliary water inlet cap (for endoscopes with auxiliary water feeding only)” on page 49. Auxiliary water inlet Auxiliary water inlet cap Endoscope connector Figure 3.12 EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
  • Page 50 Confirm that the mouthpiece is free from cracks, deformation or discoloration (see Figure 3.14). Using your fingers, check for excessive scratching or other irregularities on all surfaces of the mouthpiece (see Figure 3.14). EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
  • Page 51: Attaching Accessories To The Endoscope

    Attach the suction valve to the suction cylinder of the endoscope (see Figures 3.15 and 3.16). Confirm that the valve fits properly without any bulging of the skirt. Also confirm that the valve cannot be rotated. EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
  • Page 52 Confirm that the valve fits properly without any bulging of the skirt. Air/water valve Suction valve Skirt Suction cylinder Air/water cylinder Figure 3.16 The air/water valve may stick at first, but it should operate smoothly after it is depressed a few times. EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
  • Page 53 If the auxiliary water inlet cap is not attached, attach the fitting ring to the auxiliary water inlet on the endoscope connector (see Figure 3.18). Fitting ring Auxiliary water inlet cap Auxiliary water inlet Endoscope connector Figure 3.18 EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
  • Page 54: Inspection And Connection Of Ancillary Equipment

    Connect the water container’s connection adapter to the air supply connector and water supply connector (see Figure 3.19). Confirm that the water container’s connection adapter fits properly and that it cannot be rotated. EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
  • Page 55 Turn the connector of the videoscope cable clockwise until it stops (see Figure 3.20). Confirm that the mark on the videoscope cable is aligned with mark 2 on the endoscope connector. EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
  • Page 56 (for endoscopes with auxiliary water feeding only, see Figure 3.22). Auxiliary water tube Auxiliary water inlet Auxiliary water inlet cap Endoscope connector Figure 3.22 EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
  • Page 57: Inspection Of The Endoscopic System

    10 cm and confirm that no air bubbles are emitted when the air/water valve is not operated. Cover the hole in the air/water valve with your finger and confirm that air bubbles are continuously emitted from the air/water nozzle. EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
  • Page 58 Observe the endoscopic image and confirm that water flows on the entire objective lens. Release the air/water valve. Observe the endoscopic image and confirm that the emission of water stops and that the valve returns smoothly to its original position. EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
  • Page 59 Remove the distal end of the endoscope from the water. Depress the suction valve and aspirate air for a few seconds to remove any water from the instrument channel. EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
  • Page 60 Disconnect the water tube from the water pump or the syringe from the luer port of the auxiliary water tube. Make sure that no water leaks from the luer port of the auxiliary water tube and/or the distal end of the insertion tube. EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
  • Page 61 This could cause damage to and/or malfunction of the equipment. Syringe or water tube from a water pump Luer port Auxiliary water tube Connecting end Figure 3.23 EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
  • Page 62: Chapter 4 Operation

    In this case, the distal end of the endoscope can become hot and could cause operator and/or patient burns (when using a light source other than CLV-180). EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
  • Page 63 If any of the following phenomena occur during an examination, immediately stop the examination and withdraw the endoscope from the patient as described in Section 5.2, “Withdrawal of the endoscope with an abnormality” on page 83. EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
  • Page 64 (for endoscopes with flexibility adjustment only). EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
  • Page 65 If this vapor continues to interfere with the examination, remove the endoscope, wipe the distal end with a lint-free cloth moistened with 70% ethyl or isopropyl alcohol, reinsert the endoscope and continue the examination. EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
  • Page 66: Insertion

    If resistance to insertion is felt, or the patient reports pain, stop insertion immediately. Otherwise, operator and/or patient injury can result or the endoscope could become lodged and be difficult to withdraw. EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
  • Page 67 10 cm or less from the junction of the boot. Insertion tube damage can occur (see Figures 4.2 and 4.3). > 10 cm Figure 4.2 Endoscopes with flexibility adjustment > 10 cm Figure 4.3 EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
  • Page 68 Operate the angulation control knobs as necessary to guide the distal end for insertion and observation. The endoscope’s angulation locks are used to hold the angulated distal end in position. EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
  • Page 69 In this case, straighten the insertion tube as much as possible before attempting to increase the rigidity. Failure to do so may cause patient pain, injury, bleeding and/or perforation. EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
  • Page 70 EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
  • Page 71 Cover the air/water valve’s hole to feed air from the air/water nozzle at the distal end (see Figure 4.5). Depress the air/water valve to feed water onto the objective lens (see Figure 4.5). Suction valve Air/water valve Figure 4.5 EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
  • Page 72 Depress the suction valve to aspirate excess fluid or other debris obscuring the endoscopic image (see Figure 4.5). Performing both air feeding and suction at the same time sometimes makes it easier to remove water droplets from the objective lens surface. EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
  • Page 73 ∗1 NBI stands for Narrow Band Imaging. For more details, refer to the instruction manual of the CV-180. EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
  • Page 74: Using Endo-Therapy Accessories

    If the distal end of an endo-therapy accessory is not visible in the endoscopic image, do not open the distal end or extend the needle of the instrument. This could cause patient injury, bleeding, perforation and/or equipment damage. EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
  • Page 75 When not using an endo-therapy accessory, attach the cap to the main body of the biopsy valve. EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
  • Page 76 Hold the endo-therapy accessory approximately 4 cm from the biopsy valve and advance it slowly and straight into the biopsy valve using short strokes while observing the endoscopic image. EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
  • Page 77 Take care not to cause tissue trauma. Withdraw the endo-therapy accessory slowly while the tip of the endo-therapy accessory is closed and/or retracted into its sheath. EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
  • Page 78 Prepare, inspect and connect the electrosurgical unit and electrosurgical accessories as described in their instruction manuals. The application of high-frequency current may interfere with the endoscopic image. This does not indicate a malfunction. EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
  • Page 79 Do not use a damaged laser probe. A laser probe with a damaged sheath or distal end may cause patient injury and/or equipment damage. Prepare, inspect and connect the laser unit and laser probe as described in their instruction manuals. EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
  • Page 80: Withdrawal Of The Endoscope

    (for CF/PCF models only). Carefully withdraw the endoscope while observing the endoscopic image. Remove the mouthpiece from the patient’s mouth (for GIF models only). EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
  • Page 81: Transportation Of The Endoscope

    (see Figure 4.7). Figure 4.7 EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
  • Page 82 Putting the endoscope in the carrying case while the insertion tube is rigid could damage the endoscope (for endoscopes with flexibility adjustment only). EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
  • Page 83: Chapter 5 Troubleshooting

    Some problems that appear to be malfunctions may be correctable by referring to Section 5.1, “Troubleshooting guide”. If the problem cannot be resolved by the described remedial action, stop using the endoscope and send it to Olympus for repair. Olympus does not repair accessory parts. If an accessory part becomes damaged, contact Olympus to purchase a replacement.
  • Page 84 Replace it with a new one. damaged. The air/water An incorrect air/water Use a correct air/water valve. valve cannot be valve is used. attached. The air/water valve is Replace it with a new one. damaged. EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
  • Page 85 An image is An incompatible video Select a compatible video system center. abnormal. system center is being used. An incompatible light Select a compatible light source. source is being used. EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
  • Page 86 Operate the correct remote switch. switch does not switch is operated. work. The remote switch Set the remote switch function correctly as function has been set described in the video system center’s improperly. instruction manual. EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
  • Page 87: Withdrawal Of The Endoscope With An Abnormality

    (for CF/PCF models only). Carefully withdraw the endoscope while observing the endoscopic image. Remove the mouthpiece from the patient’s mouth (for GIF models only). EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
  • Page 88 (for CF/PCF models only). Carefully withdraw the endoscope while observing the endoscopic image. Remove the mouthpiece from the patient’s mouth (for GIF models only). EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
  • Page 89 (for CF/PCF models only). Release the angulation control knobs and carefully withdraw the endoscope. Remove the mouthpiece from the patient’s mouth (for GIF models only). EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
  • Page 90: Returning The Endoscope For Repair

    Olympus. Before returning the endoscope for repair, contact Olympus. With the endoscope, include a description of the malfunction or damage and the name and telephone number of the individual at your location who is most familiar with the problem.
  • Page 91: Appendix

    For further details, contact Olympus. If combinations of equipment other than those shown below are used, the full responsibility is assumed by the medical treatment facility. EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
  • Page 92 EVIS video system center ∗ ∗ center (CV-160 EVIS EXERA II video system (CV-140 center (CV-180) EVIS universal light source EVIS EXERA light source EVIS EXERA II light source (CLV-U40) (CLV-160) (CLV-180) EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
  • Page 93 (for CF/PCF models only) Sclerotherapy tube (for GIF models only) ∗ Not compatible with GIF-N180 Sclerotherapy balloon ∗ Use a non-flammable gas. (for GIF models except GIF-N180) ∗ Not compatible with GIF-H180 and CF-H180AL/I EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
  • Page 94 Water-resistant (BW-201T) (MAJ-1339) cap (MH-553) Channel-opening Injection tube Suction cleaning cleaning brush (MH-946) adapter (MH-856) (MH-507) Chain for water-resistant cap (MAJ-1119) ∗ These products may not be available in some areas. EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
  • Page 95 Accessories (for CF/PCF models only) Splinting tube ∗1 ST-C3 ST-C3S ST-C5 Endoscope ST-C8 CF-Q180AL/I – CF-H180AL/I – PCF-Q180AL/I – – – applicable – not applicable ∗1 This accessory may not be available in some areas. EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
  • Page 96 FB-36K-1 CF-Q180AL FB-24U-1 FB-50U-1 FB-37U-1 – CF-Q180AI FB-24Q-1 FB-50Q-1 FB-37U-1 – CF-H180AL FB-24U-1 FB-50U-1 FB-37U-1 – CF-H180AI FB-24Q-1 FB-50Q-1 FB-37U-1 – PCF-Q180AL FB-24U-1 – FB-37U-1 – PCF-Q180AI FB-24Q-1 – FB-37U-1 – EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
  • Page 97 FB-25KR-1 CF-Q180AL FB-53U-1 FB-54U-1 FB-55U-1 – CF-Q180AI FB-53Q-1 FB-54Q-1 FB-55Q-1 – CF-H180AL FB-53U-1 FB-54U-1 FB-55U-1 – CF-H180AI FB-53Q-1 FB-54Q-1 FB-55Q-1 – PCF-Q180AL FB-53U-1 FB-54U-1 FB-55U-1 – PCF-Q180AI FB-53Q-1 FB-54Q-1 FB-55Q-1 – EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
  • Page 98 FB-240K CF-Q180AL FB-212U FB-222U FB-232U FB-242U CF-Q180AI FB-212U FB-222U FB-232U FB-242U CF-H180AL FB-212U FB-222U FB-232U FB-242U CF-H180AI FB-212U FB-222U FB-232U FB-242U PCF-Q180AL FB-210U FB-220U FB-230U FB-240U PCF-Q180AI FB-210U FB-220U FB-230U FB-240U EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
  • Page 99 FG-49L-1 CF-Q180AL – – FG-16U-1 FG-45U-1 CF-Q180AI – – FG-16U-1 FG-45U-1 CF-H180AL – – FG-16U-1 FG-45U-1 CF-H180AI – – FG-16U-1 FG-45U-1 PCF-Q180AL – – FG-16U-1 FG-45U-1 PCF-Q180AI – – FG-16U-1 FG-45U-1 EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
  • Page 100 Standard type Spray type Endoscope GIF-N180 PW-2L-1 PW-6P-1 GIF-Q180 PW-1L-1 PW-5L-1 GIF-H180 PW-1L-1 PW-5L-1 CF-Q180AL PW-1V-1 PW-5V-1 CF-Q180AI PW-1V-1 PW-5V-1 CF-H180AL PW-1V-1 PW-5V-1 CF-H180AI PW-1V-1 PW-5V-1 PCF-Q180AL PW-1V-1 PW-5V-1 PCF-Q180AI PW-1V-1 PW-5V-1 EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
  • Page 101 NM-200U CD-11Z, CD-110U CF-H180AI HX-400U-30 NM-4U-1 NM-200U CD-11Z, CD-110U PCF-Q180AL HX-400U-30 NM-4U-1 NM-200U CD-21Z, CD-120U PCF-Q180AI HX-400U-30 NM-4U-1 NM-200U CD-21Z, CD-120U ∗1 These accessories may not be available in some areas. EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
  • Page 102 D-201-14304 CF-Q180AI – – D-201-14304 CF-H180AL – – D-201-15004 CF-H180AI – – D-201-15004 PCF-Q180AL – MAJ-297 D-201-12704 PCF-Q180AI – MAJ-297 D-201-12704 ∗1 These accessories may not be available in some areas. EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
  • Page 103 SD-210U-15 CF-Q180AL SD-16U-1 SD-17U-1 SD-210U-25 SD-210U-15 CF-Q180AI SD-16U-1 SD-17U-1 SD-210U-25 SD-210U-15 CF-H180AL SD-16U-1 SD-17U-1 SD-210U-25 SD-210U-15 CF-H180AI SD-16U-1 SD-17U-1 SD-210U-25 SD-210U-15 PCF-Q180AL SD-16U-1 SD-17U-1 SD-210U-25 SD-210U-15 PCF-Q180AI SD-16U-1 SD-17U-1 SD-210U-25 SD-210U-15 EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
  • Page 104 KD-1L-1 CF-Q180AL FD-2U-1 FD-210U FD-230U – CF-Q180AI FD-2U-1 FD-210U FD-230U – CF-H180AL FD-2U-1 FD-210U FD-230U – CF-H180AI FD-2U-1 FD-210U FD-230U – PCF-Q180AL FD-1U-1 FD-210U FD-230U – PCF-Q180AI FD-1U-1 FD-210U FD-230U – EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
  • Page 105: Emc Information

    Voltage Complies This instrument stabilizes its own radio variability and has no affect fluctuations/flicker such as flicker in lighting apparatus. emissions IEC 61000-3-3 EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
  • Page 106 It is recommended to use this instrument by (50/60 Hz) maintaining enough distance from any magnetic field equipment that operates with high current. IEC 61000-4-8 is the AC mains power supply prior to application of the test level. EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
  • Page 107 • Electromagnetic interference may occur on this instrument near a high-frequency electrosurgical equipment and/or other equipment marked with the following symbol: EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
  • Page 108 EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL...
  • Page 109 ©2005 OLYMPUS MEDICAL SYSTEMS CORP. All rights reserved. No part of this publication may be reproduced or distributed without the express written permission of OLYMPUS MEDICAL SYSTEMS CORP. OLYMPUS is a registered trademark of OLYMPUS CORPORATION. Trademarks, product names, logos, or trade names used in this document are generally registered trademarks or trademarks of each company.
  • Page 110 Manufactured by 2951 Ishikawa-cho, Hachioji-shi, Tokyo 192-8507, Japan Fax: (042)646-2429 Telephone: (042)642-2111 Distributed by 3500 Corporate Parkway, P.O. Box 610 Center Valley, PA 18034-0610, U.S.A. Fax: (484)896-7128 Telephone: (484)896-5000 One Corporate Drive, Orangeburg, N.Y. 10962, U.S.A. Fax: (845)398-9444 Telephone: (845)398-9400 5301 Blue Lagoon Drive, Suite 290 Miami, FL 33126-2097, U.S.A.