Chapter 3 Preparation and Inspection
3.5
Inspection and connection of ancillary
equipment
Inspection of ancillary equipment
Connection of the endoscope and ancillary equipment
50
Attach the water container to the specified receptacle on the
trolley or the light source. If the water container is attached
anywhere else, water may drip from the water container's
water supply tube, and equipment malfunction can result.
Take care not to spill water from the water container's
connection adapter when detaching the connection adapter
from the endoscope. Spilled water could splash on the
equipment, and it may cause equipment malfunction.
Prepare and inspect the light source, video system center, video monitor, water
container, suction pump and endo-therapy accessories as described in their
respective instruction manuals.
Firmly connect the suction tube from the suction pump to the
suction connector on the endoscope connector. If the suction
tube is not attached properly, debris may drip from the tube
and can present an infection-control risk, damage and/or
reduce suction capability.
The CV-100 is not compatible with the GIF-N180. If the
GIF-N180 is used with the CV-100, the endoscopic image
may not appear on the video monitor.
The GIF-H180 and CF-H180AL/I can only be connected to
the CV-180.
1.
If any ancillary equipment is ON, turn it OFF.
2.
Insert the endoscope connector completely into the scope socket (output
socket when using the CLV-U40) of the light source.
3.
Connect the water container's connection adapter to the air supply
connector and water supply connector (see Figure 3.19).
4.
Confirm that the water container's connection adapter fits properly and that
it cannot be rotated.
EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL