User Manual for the Thermographic System VarioCAM
9 Operation
Status "Spoteps"
In the "Spoteps" status, the emissivity of a measuring area can be changed from 0.01 to 1.00.
Fig. 49
View of the spot editor in the "Spotdim" status
The emissivity of the respectively active ROI can be changed using the buttons. When pressed briefly,
buttons ,
respectively reduce or increase emissivity by 0.1; buttons ,
increase emissivity by 0.01. When pressing the buttons for a longer time, emmissivity may be adjusted
faster by larger increments.
When generating an ROI, the globally valid emissivity is, first of all, adopted for this ROI (sea 9.6.3 Menu
"Measure", page 33). The emissivity is to be changed in Spoteps for each ROI individually. Differences
between local emissivity and the globally set value are shown by changes in the font colour of the
identifier of the respective ROI: It will then be purple (instead of black if global = local emissivity).
Emissivity settings will affect the temperature calculation within a measuring area.
User Manual
®
hr head
Spot editor: "Spoteps"
Help for buttons movements for
changing the emissivity in
0.01/0.1 increments
Active ROI 2
Emissivity of ROI 2
© InfraTec GmbH 2015
respectively reduce or
19