Preparing the Pipe
1. Pick a mounting location with the longest straight run. You must have easy access to at
2. Decide on your mounting mode (Direct or Reflect). Always use Reflect Mode whenever
3. After receiving the spacing dimensions from the Installation Menu, prepare the pipe
4. Clean an area 13 mm (1/2-inch) on either side of the sensors.
5. Clean an additional 13 mm (1/2-inch) along the length of the sensors.
FUG1010 IP65 NEMA 4X
Operating Instructions, 01/2013, A5E02951519-AC
least one side of your pipe. The pipe at the mounting location must remain full, even at
zero flow.
possible. You may only need to use Direct Mode if your pipe is plastic.
surface. De-grease the surface, if necessary, and remove any grit, corrosion, rust, loose
paint, etc. Use abrasive material provided to provide a clean contact surface for the
sensors.
Note
Please note that the instructions show vertical mounting for clarity purposes only. Do not
install sensors on the top of a pipe.
Pipe
Cleaned Areas
Figure 5-9
Pipe Surface Preparation
Connecting
5.5 Sensor Installation