7.13 Resistive Temperature Device (RTD) Calibration
5. To enable numeric entry <Right Arrow>, then type in the reading of the reference
6. To recalibrate the RTD sensor <ENTER>. To verify the calibrated reading, go to the Dual
Note
Factory Calibration provides an additional prompt after a new temperature is entered:
[Are you Sure? No Yes]. It is recommended that you use [User Cal] to avoid alteration of
preset factory calibration.
Ice Bath RTD Calibration
Use deionized water and ice mixture at 0°C (32°F) equilibrium for an ice bath. Ensure
temperature with a reference thermometer. Siemens can not assume responsibility for the
incorrect design, construction or operation of an Ice Bath.
NOTICE
Sensor Damage
If RTD sensor makes direct contact with ice during an ice bath calibration procedure the
sensor may be damaged and the calibration results will be incorrect.
Do not allow an RTD sensor to make direct contact with ice during an ice bath calibration
procedure.
To perform a 0°C (32°F) calibration:
1. Immerse RTD sensor in deionized water and ice mixture. Stir the mixture constantly.
2. In the [RTD Calibrate] menu move the highlight by the pressing <Up/Down Arrow> to the
3. To access the RTD option list press <Right Arrow>. Move the highlight to [User Cal] then
4. After the RTD sensor reaches equilibrium at 0°C (32°F), press <ENTER> to recalibrate
5. To verify the calibrated reading, go to Dual Path Flow [Diagnostic Data/Gas Data] menu
136
thermometer (e.g., 72.0).
Path Flow menu [Diagnostic Data/Gas Data] menu to check the current RTD output.
Make sure that it coincides with the gas reading of the reference thermometer. Repeat for
the other RTD, if necessary.
RTD you want to calibrate (RTD 1 or RTD 2).
press <ENTER>. This triggers the pop-up window.
the RTD sensor.
to check the current RTD output. Make sure that it coincides with the gas reading of the
reference thermometer. Repeat for the other RTD, if necessary.