Siemens FUG1010 IP65 NEMA 4X Operating Instructions Manual page 133

Sitrans f, ultrasonic flowmeters
Hide thumbs Also See for FUG1010 IP65 NEMA 4X:
Table of Contents

Advertisement

Functions
7.11 Setting Relays
S Fltwarn
S Soft Fault
4 Fault Alarm
4 Spacing
4 Turbulence
3 Fault Alarm
3 Spacing
3 Turbulence
2 Fault Alarm
2 Spacing
2 Turbulence
1 Fault Alarm
1 Spacing
1 Turbulence
Note
For Multi-Path flow meters: "1" = Path 1, "2" = Path 2, "3" = Path 3, "4" = Path 4 and "S"
represents the system or average channel. These characters appear to the left of the option
list parameter.
Assigning functions to Relay 1:
1. From the [Dual Path Flow] menu scroll down and highlight [I/O Data Control].
2. Press <Right Arrow> and scroll down to [Relay Setup].
3. To access the [Relay Setup] option list press <Right Arrow>.
132
Fault warning occurs when 1 or more paths are in fault.
Fault condition - Memory mode active.
Loses receive signal (all paths in fault).
Sensor spacing needs adjusting.
Turbulence percentage exceeds alarm setpoint.
Loses receive signal (all paths in fault).
Sensor spacing needs adjusting.
Turbulence percentage exceeds alarm setpoint.
Loses receive signal (all paths in fault).
Sensor spacing needs adjusting.
Turbulence percentage exceeds alarm setpoint.
Loses receive signal (all paths in fault).
Sensor spacing needs adjusting.
Turbulence percentage exceeds alarm setpoint.
Operating Instructions, 01/2013, A5E02951519-AC