Siemens FUG1010 IP65 NEMA 4X Operating Instructions Manual page 164

Sitrans f, ultrasonic flowmeters
Hide thumbs Also See for FUG1010 IP65 NEMA 4X:
Table of Contents

Advertisement

2. A typical menu screen will appear as shown below and indicate the current ALC (e.g.,
3. To exit Force Transmit, press the <Left Arrow> and a Detection Fault prompt will appear
4. Press the <Left Arrow> again and the meter will return to the [Install Sensor] menu and
Setting a Forced Frequency
1. To force a frequency, repeat steps 1 and 2 above, but press <Right Arrow>. The following
2. Using numeric keys enter the frequency and press <ENTER>.
3. To complete the Install process after mounting the transducers press <ENTER>.
4. If the Force Transmit diagnostic procedure is not used, the normal [Install Complete]
FUG1010 IP65 NEMA 4X
Operating Instructions, 01/2013, A5E02951519-AC
50). This ALC number indicates the current receive signal strength and can be used for
further diagnostic purposes.
(see above).
highlight the [Empty Pipe Set] menu cell.
typical display line will appear: Drive =0
function occurs.
Troubleshooting
10.5 Force Transmit
163