Honeywell DR4300 Product Manual page 99

Circular chart recorder
Hide thumbs Also See for DR4300:
Table of Contents

Advertisement

Lower Display
Prompt
LD CAL
FAC [factory setting]
FLD
INP HI
–999 to 9999
[factory setting = 900]
–999 to 9999
INP LO
in Engineering Units
[factory setting =
− 300]
BIAS
–99.9 to 999.9
[factory setting = 0]
0 to 120 seconds
FILTER
No filter = 0
[factory setting = 0]
4/00
Upper Display
Range of Setting
or Selection
LOAD CALIBRATION VALUES—This parameter is used to specify
which set of calibration values you want to use.
Factory—Uses built-in calibration values (stored in read-only
memory).
Field—Uses calibration values stored as a result of the field
calibration procedure described in Section 5.
ATTENTION
FLD uses the same calibration values as FAC. When field
calibration is performed, the value of LD CAL is automatically set to
FLD at the completion of the field calibration operation.
INPUT HIGH RANGE VALUE in engineering units is displayed but
can only be configured for linear input type.
Otherwise, this is a read-only display of the higher range value for
the selected T/C or RTD input type.
• Scale the input signal to the display value you want for 100 %.
• EXAMPLE:
Actuation (Input) = 4 to 20 mA
Process Variable = Flow
Range of Flow = 0 to 250 Gal/Min
High Range display value = 250
Then 20 mA = 250 Gal/Min
INPUT LOW RANGE VALUE in engineering units is displayed but
can only be configured for linear input type.
• Otherwise, this is a read-only display of the low range value for
the selected T/C or RTD input type.
• Scale the input signal to the display value you want for 0 %. See
example in "INPHI".
INPUT BIAS—Used to compensate the input for drift of an input
value due to deterioration of a sensor, or some other cause; select
the value you want on the input.
FILTER FOR INPUT—A software digital filter is provided for the
input to smooth the input signal. You can configure the first order
lag time constant from 1 to 120 seconds. If you do not want
filtering, enter 0.
DR4300 Circular Chart Recorder Product Manual
Configuration, Startup, and Operation
of Recorder with Display
Parameter
Definition
Until field calibration has been done,
87