Step
Description
4
Exit Calibration Mode
4/00
Press
FUNC
DISP
or
SET
UP
DR4300 Circular Chart Recorder Product Manual
Input and Output Calibration for Recorder with Display
Action/Result
The recorder stores the span value.
Repeat procedure for the other pen. Remove power from
recorder and disconnect the milliammeter and connect
output wiring.
To exit the calibration mode.