Honeywell DR4300 Product Manual page 35

Circular chart recorder
Hide thumbs Also See for DR4300:
Table of Contents

Advertisement

Taking electrical noise precautions
Electrical noise is composed of unabated electrical signals which produce undesirable effects in
measurement and control circuits.
Digital equipment is especially sensitive to the effects of electrical noise. Your recorder has built-in circuits
to reduce the effect of electrical noise from various sources. If there is a need to further reduce these effects:
• Separate External Wiring—Separate connecting wires into bundles (see Table 2-6) and route the
individual bundles through separate conduits or metal trays.
• Use Suppression Devices—For additional noise protection, you may want to add suppression devices
at the external source. Appropriate suppression devices are commercially available.
For additional noise information, refer to document 51-52-05-01, How to Apply Digital Instrumentation in
Severe Electrical Noise Environments.
CE conformity special conditions (Europe)
Shielded cables with a drain wire are required for all input and output cables. All instrument shielded cable
drain wires must be connected to a low impedance earth ground at the entry fitting. Cable/conduit entry
fittings shall effectively terminate the cable shield connecting the shield to the enclosure conductive coating.
The shielded cable shield shall not extend into the instrument closure.
3/00
DR4300 Circular Chart Recorder Product Manual
Installation
23

Advertisement

Table of Contents
loading

This manual is also suitable for:

Dr4300 seriesDr4322Dr4321Dr4301Dr4302Dr4311 ... Show all

Table of Contents