VM700T Option 40/41/42 Audio Measurements User Manual
NOTE. Use a test signal level of at least –60 dBu and test at least 10 frequencies
over the entire range. Tested frequencies should include the specification limits
(20 Hz, 20 kHz) and the 40 Hz breakpoint. Example frequencies to test might
also include the following: 100 Hz, 200 Hz, 1 kHz, 2 kHz, 10 kHz.
3. Check that the measured frequency is accurate to the limits shown in
Table B-2 for the applied test signal frequency and level and that the phase
difference between channels meets the phase difference specification.
Appendix B: Verification Procedure
B-7