RIGOL
Return Format:
The query returns 5.000e-004, and the unit is s.
18. :MEASure:PDUTycycle?
Command Format:
:MEASure:PDUTycycle?□[<source>]
Function:
The command measures the positive duty cycle of signal from <source>. The
<source> may be CHANnel1, CHANnel2, CHANnel3 or CHANnel4.
Return Format:
The query returns 5.000e001, and the unit is %.
19. :MEASure:NDUTycycle?
Command Format:
:MEASure:NDUTycycle?□[<source>]
Function:
The command measures the negative duty cycle of signal from <source>. The
<source> may be CHANnel1, CHANnel2, CHANnel3 or CHANnel4.
Return Format:
The query returns 5.000e001, and the unit is %.
20. :MEASure:PDELay?
Command Format:
:MEASure:PDELay?□[<source A>,<source B>]
Function:
The command measures the delay between <sourceA> and <sourceB> relative
to the rising edge. The <source> may be CHANnel1, CHANnel2, CHANnel3 or
CHANnel4.
Return Format:
The query returns <-1.000 e-004, and the unit is s.
2-28