Eclipse Additional Information
Right clicking during testing
cally when possible. The red area indicates the intensities that cannot be started right away.
Waiting frequencies also depend on the intensity level of other frequencies being tested for the same ear.
The difference between the neighbouring frequencies can cause stimulation on the basilar membrane on a
neighbour frequency (i.e. if the selected intensity is 10dB nHL at 1000Hz and 70dB nHL at 500Hz). To avoid
the risk of stimulating neighbour frequencies the system limits the frequency intensity that can run at the
same time per ear to a 20 dB range.
Setting new intensity range will take place when the new pending or selected intensity needs to change
the attenuator. This should take place within a few seconds and once set the test will automatically continue.

6.4.2.4 Right clicking on a handle

Right clicking with the mouse on any curve handle provides these options:
Delete will delete this particular curve.
Delete All will delete all curves for this frequency in this ear.
Hide will hide the actual curve, but leave the handle of the curve on the screen to indicate that this intensity
has been tested and is available for display if needed. When right clicking on a hidden curve Show will ap-
pear from the dialog box.
Cursor brings forward a cursor that will track with the curve and provide numerical information for the proba-
bility percentage and time at any point of the curve.
Right clicking with the mouse in this area will provide
access to control of the stimulus for that specific au-
diogram point (e.g. 4000Hz, Left ear).
All 8 audiogram points can be controlled individually
by right clicking in their specific areas.
Right click on the desired frequency To extend test
time, to Stop an ongoing intensity immediately or
to Add the next test intensity for the selected fre-
quency.
Detected and non-detected responses are indicated
by green and red color dots while the test is running.
This is helpful to determine the next intensity to be
initiated.
Even when an intensity is running additional intensi-
ties can be added to the queue. The added intensi-
ties will be set to waiting, which is indicated in the
window when right clicking, and will start automati-
Page 135