Eclipse Additional Information
Page 107
After the marker is selected, place it on the desired peak or trough.
A second way of marking the peak from the Edit tab is to double click the waveform handle (to select it) you
would like to mark. Choose the desired marker on the left side of sheet and drag your mouse to the correct
point and click to place the marker. You can also choose the shortcut keys 1-4 on the keyboard to bring up
the appropriate marker, and then move by the mouse of arrow keys. Finally, hit Enter to set. The N1 and P1
markers are the ones used for calculating the asymmetry ratio.
2. Setting up L & R Waveform VEMP Partners
After collection, choose a left or right ear by double clicking the waveform handle.
Next, right click the waveform handle of the opposite ear and select Set as VEMP Partner.
These two selected waveforms will be used in the Asymmetry Calculation.
NOTE The VEMP asymmetry ratio is not calculated before a VEMP partner is selected.
NOTE Setting VEMP partners is only possible when right and left tests have been performed with same stim-
ulus and intensity.
Need help?
Do you have a question about the Eclipse and is the answer not in the manual?
Questions and answers