Eclipse Additional Information

6.4.2 The ASSR tab

From the ASSR tab you can monitor and control your test. This is the screen where you see the resulting es-
timated audiogram and where you apply whatever corrections are needed.

6.4.2.1 The left panel user interface

The left panel user interface is available from both the ASSR tab and the Audiogram tab.
The left panel contains information about each frequency and the intensity that is currently running and
which intensities that are waiting. The information is split into left and right ear separately.
The Total Session Status indicates the total test time and the used transducer (e.g. In-
sert phone).
The Stimulus rate is a dropdown menu where the stimulus rate can be set before test-
ing, or the user can change during the testing by pressing stop and restarting the test.
The waiting status indicates the order of next intensity to be enabled.
The Selected Correction Factor is only available in the Audiogram Tab and allows you
to select the correction factors for the measurement.
The Start/Stop button is used to start data acquisition. The button is dimmed if no test
can be carried out (e.g. no ECLIPSE is connected, or no test stimuli are selected).
The Pause button allows you to stop the data acquisition temporarily, e.g. if the patient
is becoming restless.
Page 133