Battery Installation - Cardiac Science POWER2HEART AED G3 Automatic 9300A Operation And Service Manual

Automated external defibrillators
Table of Contents

Advertisement

BATTERY SHELF LIFE
The Cardiac Science AED batteries have a shelf life of five years. Shelf life is defined as the length of time a
battery can be stored, prior to installation into AED, without degrading its performance.
Note: Storing the battery outside its specific range (0-50°C) will decrease battery life.

BATTERY INSTALLATION

1. With the label on the battery facing the AED battery compartment, insert the battery as
shown in the drawing.
2. Push the latched end of the battery firmly into the AED, as shown in the drawing, until the
battery snaps into place. The exposed side of the battery should be flush with the outside of
the AED case.
3. For the Powerheart AED, open the lid for 5 seconds to initiate self-test. If the battery is
installed properly, the STATUS INDICATOR will turn GREEN . Close the lid.
OR
Open the lid for 5 seconds to initiate a self-test. If the battery is installed properly, the
SMARTGAUGE battery indicator LEDs will illuminate; additionally, on the Powerheart AED, the
STATUS INDICATOR will turn GREEN . If service is required, then the SERVICE indicator will
illuminate; call for service.
Note: Battery part number 9146 is only for use with the Powerheart AED G3 and
Powerheart AED G3 Automatic. Battery part number 9143 is only for use with the FirstSave AED G3.
WARNING: Battery is Not Rechargeable
Do not attempt to recharge the battery. Any attempt to recharge the battery may result in an
explosion or fire hazard.
CAUTION: Lithium Sulfur Dioxide Battery
Pressurized contents: never recharge, short circuit, puncture, deform, or expose to temperatures
above 65°C (149°F). Remove the battery when discharged.
CAUTION: Battery Disposal
Recycle or dispose of the lithium battery in accordance with all federal, state and local laws. To
avoid fire and explosion hazard, do not burn or incinerate the battery.
CAUTION: Use only Cardiac Science Approved Equipment
Using batteries, pads, cables, or optional equipment other than those approved by
Cardiac Science may cause the Powerheart AED to function improperly during a rescue.
CAUTION: Possible Improper AED Performance
Using pads that are damaged or expired may result in improper AED performance.
© 2003-2004 Cardiac Science, Inc. All rights reserved.
Section 3 - Getting Started
112-2025-005 Rev B
page 25

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Power2heart aed g3 9300eFirstsave aed g3 9300cPowerheart aed g39300aFirstsave aed g39300c ... Show all

Table of Contents