These protocols are selected by using our MDLink software program. The five biphasic energy protocols
available are as follows:
Energy
Protocols
Factory Default
Protocol #2
Protocol #3
Protocol #4
Protocol #5
The FirstSave AED will deliver Protocol 5 only, which is variable energy 200J.
1
The ultra-low current, low current and high current shocks are variable energy. The actual energy is determined by the patient's impedance.
© 2003-2004 Cardiac Science, Inc. All rights reserved.
Shock
Energy
Sequence
Level
1
1.
200VE
2.
300VE
3.
300VE
1.
200VE
2.
200VE
3.
300VE
1.
150VE
2.
200VE
3.
200VE
1.
150VE
2.
150VE
3.
200VE
1.
200VE
2.
200VE
3.
200VE
Section 2 - Introduction
Energy Range (J)
140J-250J
190J-360J
190J-360J
140J-250J
140J-250J
190J-360J
105J-195J
140J-250J
140J-250J
105J-195J
105J-195J
140J-250J
140J-250J
140J-250J
140J-250J
112-2025-005 Rev B
page 19