AED INDICATORS

The following indicators are located on the AED.
RESCUEREADY STATUS INDICATOR FOR POWERHEART AED ONLY
The STATUS INDICATOR is located on the Powerheart AED handle. When this indicator
is GREEN , the Powerheart AED is RescueReady. This means the Powerheart AED
self-tests have verified the following:
Battery has an adequate charge.
Pads are properly connected to the Powerheart AED and in working order.
Integrity of the internal circuitry is good.
When the STATUS INDICATOR is RED , maintenance is required.
AUDIBLE MAINTENANCE INDICATOR
When the daily, weekly or monthly self-test determines service is required, an audible beep is sounded every
30 seconds until the lid is opened or the battery power is depleted. Opening and closing the lid may deactivate
the beep. If the error is not corrected by the next automatic self-test, the beep will be reactivated.