PADS
•
Self-adhesive, disposable defibrillation pads
•
Minimum combined surface area: 228cm
•
Extended length of lead wire: 1.3m
LITHIUM BATTERY SPECIFICATIONS
•
Output voltage: 12VDC (max)
•
Batteries are non-rechargeable
•
Lithium contents: 9.2g (max)
•
Check local regulations for disposal information
Estimated Shelf Life
Model
(from date of manufacture)
9146 Lithium
5 Years
9143 Lithium
5 Years
The battery operating life depends on the type of battery (9146 for Powerheart AED G3 or 9143 for FirstSave
AED G3), actual usage and environmental factors.
BATTERIES AND CAPACITOR CHARGE TIMES
A new battery typically takes 10 seconds to charge the AED to maximum energy.
A battery with reduced capacity causes the red LED light to initially turn ON and typically takes 13 seconds to
charge a fully discharged AED to maximum energy.
The maximum time from "Power On" to "Ready to Shock" is 28 seconds for a new rescue.
The maximum time from "Analyze" to "Ready to Shock" is 22 seconds for a new rescue.
DELIVERY OF THREE DEFIBRILLATION SHOCKS
55 seconds (nominal)
© 2003-2004 Cardiac Science, Inc. All rights reserved.
Section 7 - Technical Data
2
Full Operational
Replacement Guarantee
(from date of installation)
4 Years
1 Year
112-2025-005 Rev B
Typical Shocks
up to 290 shocks
up to 100 shocks