Cardiac Science POWER2HEART AED G3 Automatic 9300A Operation And Service Manual page 50

Automated external defibrillators
Table of Contents

Advertisement

PADS
Self-adhesive, disposable defibrillation pads
Minimum combined surface area: 228cm
Extended length of lead wire: 1.3m
LITHIUM BATTERY SPECIFICATIONS
Output voltage: 12VDC (max)
Batteries are non-rechargeable
Lithium contents: 9.2g (max)
Check local regulations for disposal information
Estimated Shelf Life
Model
(from date of manufacture)
9146 Lithium
5 Years
9143 Lithium
5 Years
The battery operating life depends on the type of battery (9146 for Powerheart AED G3 or 9143 for FirstSave
AED G3), actual usage and environmental factors.
BATTERIES AND CAPACITOR CHARGE TIMES
A new battery typically takes 10 seconds to charge the AED to maximum energy.
A battery with reduced capacity causes the red LED light to initially turn ON and typically takes 13 seconds to
charge a fully discharged AED to maximum energy.
The maximum time from "Power On" to "Ready to Shock" is 28 seconds for a new rescue.
The maximum time from "Analyze" to "Ready to Shock" is 22 seconds for a new rescue.
DELIVERY OF THREE DEFIBRILLATION SHOCKS
55 seconds (nominal)
© 2003-2004 Cardiac Science, Inc. All rights reserved.
Section 7 - Technical Data
2
Full Operational
Replacement Guarantee
(from date of installation)
4 Years
1 Year
112-2025-005 Rev B
Typical Shocks
up to 290 shocks
up to 100 shocks
page 49

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Power2heart aed g3 9300eFirstsave aed g3 9300cPowerheart aed g39300aFirstsave aed g39300c ... Show all