Cardiac Science POWER2HEART AED G3 Automatic 9300A Operation And Service Manual page 47

Automated external defibrillators
Table of Contents

Advertisement

8.
Q:
When I open the lid, why do I get the voice prompt "Data in Memory. Push Flashing Button to Erase
Data and Perform Rescue." How to I get the message to stop occurring?
A:
This message occurs when there is a previously stored rescue in the internal memory of the AED.
You can clear the message by:
1.
Pressing the SHOCK/CONTINUE button to erase the internally stored rescue data OR
2.
Retrieving the rescue data with RescueLink and erasing the stored data with RescueLink.
9.
Q:
The AED did not sound an audible alert when I removed the pads and closed the lid. Why?
A:
The lid-closed pad self-test only activates the STATUS INDICATOR (on the Powerheart AED).
The AED allows time for replacement of the pads – as removing pads is a normal
procedure after a rescue – or a battery during the post rescue procedure, however, an audible
maintenance indicator will be triggered after the next Daily Self-test.
10. Q:
What if I have to perform a rescue in an isolated area and at subzero temperatures?
A:
When travel to a rescue involves exposing the AED to extremely cold temperatures for an extended
period of time, keep the pads and the battery warm.
page 46
112-2025-005 Rev B
© 2003-2004 Cardiac Science, Inc. All rights reserved.

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the POWER2HEART AED G3 Automatic 9300A and is the answer not in the manual?

Questions and answers

This manual is also suitable for:

Power2heart aed g3 9300eFirstsave aed g3 9300cPowerheart aed g39300aFirstsave aed g39300c ... Show all

Table of Contents