VOICE PROMPT
"Stand Clear! Push
Flashing Button to
Deliver Shock."
"Check for Breathing. If Not
Breathing, Give Patient Two
Breaths."
"Check for Signs of
Circulation. If No
Circulation Start CPR."
"Check Pads"
"Battery Low"
"Analysis Interrupted. Stop
Patient Motion. "
"Open Lid to Continue Rescue"
"Rhythm Changed.
Shock Cancelled."
"Data in memory. Push
Flashing Button to Erase
Data and Perform Rescue."
"Continue CPR"
© 2003-2004 Cardiac Science, Inc. All rights reserved.
TEXT DISPLAY
STAND CLEAR
PUSH BUTTON TO SHOCK
IF NOT BREATHING GIVE
TWO BREATHS
IF NO CIRCULATION
START CPR
CHECK PADS
BATTERY LOW
ANALYSIS INTERRUPTED
STOP PATIENT MOTION
OPEN LID TO
CONTINUE RESCUE
RHYTHM CHANGED.
SHOCK CANCELLED
DATA IN MEMORY
PUSH BUTTON TO ERASE
CONTINUE CPR
Section 3 - Getting Started
SITUATION
After the AED is fully charged and ready to deliver the
defibrillation shock. The RED SHOCK indicator flashes
and the phrase repeats for 30 seconds or until the
SHOCK button is pushed.
• After the AED delivers 3 consecutive defibrillation
shocks.
• After the AED detects a non-shockable cardiac rhythm.
• When 2 1/2 minutes or more has elapsed since CPR
was last administered.
• Note: This prompt may be disabled in MDLink, for
Powerheart AED only.
• After previous prompt or if previous prompt is disabled.
• After the AED delivers 3 consecutive defibrillation
shocks.
• After the AED detects a non-shockable cardiac rhythm
during cardiac rhythm analysis.
• When 2 1/2 minutes or more has elapsed since CPR
was last administered.
Occurs when patient impedance is too low or too high.
Occurs once when the battery voltage becomes low,
although a rescue can continue for approximately 9 more
shocks. When the battery is too low to do a rescue, the
phrase repeats continuously. You must replace the
battery before continuing with the rescue. If completely
depleted, all AED activity will terminate.
When the AED detects ECG noise artifact, stop moving
or touching the patient. Remove other electronic devices
within a 5 meter radius.
When the lid is inadvertently closed during a rescue,
this prompt will repeat for 15 seconds.
When the device is prepared to shock then detects a
change in rhythm and therefore cancels the shock.
When there is a previously stored rescue in internal
memory, the yellow CONTINUE button will flash and the
phrase will repeat on the Powerheart AED G3 and
FirstSave AED G3.
During CPR mode when enabled, or when a rescue is
resumed in CPR mode after being interrupted by the
lid closing.