VOICE PROMPT
"Remove Cable to
Continue Rescue. "
"Communications Mode"
(Beep)
"Service Required"
For the Powerheart AED G3 Automatic only:
"Stand Clear! Shock will be
Delivered in 5 seconds."
"Four, three, two, one"
"Shock delivered."
"It is safe to touch the
patient."
"Data in memory. Push
Flashing Button to Erase
Data and Perform Rescue. "
page 32
TEXT DISPLAY
REMOVE CABLE
COMMUNICATIONS MODE
SERVICE REQUIRED
STAND CLEAR.
FIVE
FOUR
THREE
TWO
ONE
SHOCK DELIVERED
IT IS SAFE TO TOUCH THE
PATIENT.
DATA IN MEMORY
PUSH BUTTON TO ERASE
112-2025-005 Rev B
SITUATION
The message can be cleared by pressing the
SHOCK/CONTINUE button to erase the internally stored
rescue data or retrieving the rescue data with
RescueLink and erasing the stored data with
RescueLink. When a serial communication cable is
connected to the AED during a rescue, the phrase
repeats until the cable is disconnected.
When the lid is open and the serial communication
cable is plugged into the AED.
"One Beep" occurs in 15-second intervals during
CPR when enabled by the MDLink software program,
"Warble Beep" occurs when the AED requires
maintenance.
Occurs after the self-tests determine that the AED is not
functioning properly. The prompt "SERVICE REQUIRED "
will be heard when the lid is opened. The red SERVICE
indicator will illuminate and "SERVICE REQUIRED " will
repeat until you close the lid. After closing the lid, an
alarm beep will be heard until the battery is removed
or becomes completely depleted.
After the AED is fully charged and ready to deliver
the defibrillation shock. The SHOCK will automatically
be administered five seconds after the end of the
voice prompt.
Counts down seconds until a shock is automatically
delivered.
After a shock is automatically delivered.
Advises the rescuer when it is safe to touch the patient.
When there is a previously stored rescue in internal
memory, the blue CLEAR DATA button will flash and the
phrase will repeat.
© 2003-2004 Cardiac Science, Inc. All rights reserved.

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the POWER2HEART AED G3 Automatic 9300A and is the answer not in the manual?

Questions and answers

This manual is also suitable for:

Power2heart aed g3 9300eFirstsave aed g3 9300cPowerheart aed g39300aFirstsave aed g39300c ... Show all

Table of Contents