Set Up Template Configuration - Dräger X-dock 5300 Technical Manual

Hide thumbs Also See for X-dock 5300:
Table of Contents

Advertisement

Tests with substitute gas
No gas test, only the activated options are
-
run.
For tests with substitute gas, a switch is
Schneller
made to the test gas. So, for example, the
Begasungst
influence of temperature is reduced and a
est
greater accuracy is achieved.
Since the alarm thresholds no longer apply
here, it is always tested at a calculated
threshold. The threshold is calculated at 1/
3*test gas concentration.
For tests with substitute gas, a switch is
Erweiterter
made to the test gas. So, for example, the
Begasungst
influence of temperature is reduced and a
est
greater accuracy is achieved.
For the extended test, a check is carried out
on the test gas concentration with the set
tolerance.
For calibration with substitute gas, a switch
Calibration
is made to the test gas. So, for example, the
influence of temperature is reduced and a
greater accuracy is achieved.
Calibration uses the test gas concentration
of the test gas. Following successful
calibration, a switch is made to the
measured gas.
4. Assign the desired authorisation level.
5. Select Options if necessary.
The following options are available:
Response time test
(only for calibration)
Alarm test
Zero check
Check config.
Set config.
NOTICE
i
i
When the alarm functions are being tested, the
ambient noise must not be too loud as otherwise the
station will not be able to check the horn function.
1
(The Pac 7000 OV PAC 7000 OV-A does not support switching to
the test gas. For this reason, the X-dock performs the conversion
here. There is no compensation for the effects of temperature, so
testing/calibration is only recommended in a range of from 0-
30 °C.)
Dräger X-dock 5300/6300/6600
1
The response times of the sensors
are tested.
All the alarm functions of the gas
monitor are checked (noise, light,
vibration if configured).
The zero point is checked.
The configuration of the gas monitor
is compared with the template
activated at the station. If the
configuration does not match, an
error message is generated.
If this option is activated, the gas
monitors are configured according
to the activated template at the
station if a deviation has previously
been determined.
6. If required, select Further options:
Time sync.
Download DL
Reset DL
Reset TWA
Turn off
AutoCal
Purge
2
Generate
Print
1
Average shift values (time-weighted average) are generally
limited to eight hours exposure per day per workplace for
5 days a week during a work lifetime.
2
In the case of networked systems, Dräger recommends that
you do not activate the "Generate" option but create the
certificate using the X-dock Manager. This reduces the test
duration.
7. Save settings with OK.

4.10 Set up template configuration

NOTICE
i
i
Before
changing
recommends:
1. Transferring the new configuration template to a
gas monitor using an X-dock station.
2. Checking the new configuration on the gas monitor
using the Dräger CC-Vision PC software,
3. approving
the
distributing it to the corresponding X-dock station.
The maintenance station can configure gas monitors
automatically
using
a
configuration templates can be created using the Dräger CC-
Vision PC software. The configuration templates can be
transferred to the maintenance station using a USB memory
stick or the X-dock Manager.
1. Create a configuration template using the CC-Vision PC
software.
2. Transfer the template file to the station using a USB
memory stick or the X-dock Manager (see chapter 6.5.7 on
page 21 and online help for X-dock Manager).
3. When defining a test, select the desired option:
a. Select Check config. whether to carry out just one
check of the configuration.
The time on the gas monitor is
synchronised with the station.
If this option is selected, Reset DL
and Reset TWA are automatically
also activated to ensure that the DL
(data logger) is unique.
Download data memory and events
memory data from the gas monitor
to the station.
The data in the data and events
memories are cleared.
1
The TWA
time is reset.
The gas monitor is switched off
automatically after the test.
Perform automatic calibration (e. g.
calibrate after a faulty bump test).
The system is purged with fresh air
after every test.
Create certificate (stored in the
station as a PDF file).
Print out the certificate on a
connected USB printer.
the
configuration,
configuration
template
configuration
template.
Basics
Dräger
and
The