Dräger X-dock 5300 Technical Manual page 11

Hide thumbs Also See for X-dock 5300:
Table of Contents

Advertisement

When using a Dräger gas cylinder:
NOTICE
i
i
Upon entering the part no. of a Dräger gas cylinder, a
gas cylinder level indicator is displayed automatically,
unless this function was deactivated before (see
chapter 4.7.2 on page 12).
1. Enter the part no. of the Dräger gas cylinder.
All the necessary details for the configuration will be filled
in automatically. The batch number and the expiry date can
be entered manually in addition.
NOTICE
i
i
The values entered automatically must be matched
with those indicated on the test gas cylinder. If these
values are not identical, the value indicated on the gas
cylinder is to be considered and a manual correction of
the values must be made in the station.
2. If necessary select Further options and
gas cylinder level indicator.
3. If required, configure other gas inlets in the same way.
When using a gas cylinder from another manufacturer:
1. Create or delete test gas component.
Create a new test gas component with
Delete the current test gas component with
NOTICE
i
i
Deleting all test gas components deletes all
information of the test gas inlet.
2. Select gas.
3. Enter the gas concentration.
4. Select gas units.
5. If required, create additional test gas components.
6. The following information can optionally be given:
Part no. of the gas cylinder
Batch number of test gas cylinder
Expiry date of the gas cylinder
7. If required, enter Further options.
Dräger X-dock 5300/6300/6600
4.7.1
Further options
NOTICE
i
i
It is necessary to specify the volume, volume type,
pressure and unit to use the gas manager.
Configured test gas
components
Enter volume
Enter pressure
Hose length
, to reset the
1
Select this symbol to reset the gas cylinder level.
1. Enter volume of gas cylinder.
.
2. Specify the type of volume:
.
compr. = cylinder volume
relaxed = gas volume (cylinder volume x pressure)
3. If required, refresh test gas pressure.
4. If necessary, activate Hose > 2m if the length of the test
gas hose exceeds 2 m.
NOTICE
i
i
Dräger recommends not to exceed a hose length of
10 m for the gas supply hoses.
5. If required, configure other gas inlets in the same way.
State of gas pressure:
Display
Gas cylinder level indicator:
Display
Gas pressure Meaning
<0.4 bar
Gas pressure too low.
0.4 - 0.6 bar
Gas pressure correct.
>0.6 bar
Gas pressure too high.
Gas cylinder level
61 - 100 %)
30 - 60 %)
< 30 %
Gas cylinder level indicator deactivated
Basics
◄ Volume type
Enter unit for
pressure
◄ State of
gas pressure/
Gas cylinder level
1
indicator
11