Indicates a potentially hazardous situation which, if not avoided, could result in physical injury, or damage to the product or environment. It may also be used to alert against unsafe practices. NOTICE Indicates additional information on how to use the product. Dräger X-dock 5300/6300/6600...
Dräger recommends not exceeding a hose length of another. A system consists of one master for 3 (X-dock 5300/ 10 m for the gas feed hoses. 6300) or 6 (X-dock 6600) test gases. The X-dock 5300 5.
The charging process is not interrupted. To switch to operating mode: Tap the function key or touch the touchscreen or open or close a module flap. Menu ► ◄ Favourites bar Dräger X-dock 5300/6300/6600...
Select this button to access the previous screen. Log user in or out Select this button to log users in or out. The number in the symbol indicates the permission level (see chapter 7 on page 19). Dräger X-dock 5300/6300/6600...
> Station gas configuration. 2. If necessary select Further options and to reset the gas An overview of the test gas connections is displayed. cylinder level indicator. 3. If required, configure other test gas inlets in the same way. Dräger X-dock 5300/6300/6600...
Page 10
More 10 m for the gas feed hoses. information on this can be found in the relevant sensor data 5. If required, configure other test gas inlets in the same way. sheets. Dräger X-dock 5300/6300/6600...
The This function can be used to set whether or not a test or Test scheduler and Log-in mode test modes are configurable. calibration is carried out when a required test gas is not connected. Dräger X-dock 5300/6300/6600...
Page 12
Average shift values (time-weighted average) are generally limited to eight hours exposure per day per workplace for 5 days a week during a work lifetime. Further options ► 6. Save settings with OK. 2. Enter desired test name. Dräger X-dock 5300/6300/6600...
(max. 2 months) without a power supply must not be inserted into the station. Otherwise the as the internal buffer battery will run down.omatically once test cannot be fully evaluated correctly. the temperature is back in charging range. Dräger X-dock 5300/6300/6600...
Fast bump test including alarm testing information. Extended bump test incl. zero-point check and Remove the instrument from the module. Test 2: EXT alarm testing. Calibration, alarm test, fresh air flushing and Test 3: CAL certificate. Dräger X-dock 5300/6300/6600...
Identify and rectify the error. Repeat the test if necessary. Overview of status LEDs Colour Status Meaning blue flashing Process in progress green flashing Test passed yellow flashing Test passed, with restrictions flashing Test failed/cancelled Dräger X-dock 5300/6300/6600...
Station gas configuration System configuration Date & time Data management Language Network System address Database DBMS address Info screen Update Service interval Devices Users Station event logger Export data logger Export certificates Mobile synchronization Mass storage device Dräger X-dock 5300/6300/6600...
User profiles can be created, edited or deleted in this menu. 4. Where necessary, enter the Subnet Mask (default: Select > Data management > Users. 255.255.255.0). The user profiles will be displayed. For more information see chapter 4.6 on page 8. Dräger X-dock 5300/6300/6600...
Page 18
X-dock Manager Online Help). 1. If necessary, connect the USB storage device to the USB port on the station. 2. Select > Data management > Mobile synchronization . The available target locations are displayed. 3. Select desired target location. Dräger X-dock 5300/6300/6600...
Perform preset tests in logged-in status Change password Define test Export data memory (station) Set Station gas configuration Show devices Define extended options Manage users Set date and time Perform SW update (station) Set Language Set network configuration Database operations Dräger X-dock 5300/6300/6600...
The firmware version of the affected gas Perform an update with the CC Vision PC Module X does not support firmware version error message. monitor is not compatible with X-dock. software. Dräger X-dock 5300/6300/6600...
There must not be any older firmware files on the USB data storage device. 2. Connect the USB data storage device with firmware update to the USB port on the station. The USB icon will appear in the status bar. Dräger X-dock 5300/6300/6600...
6. Squeeze external locking lugs in and insert the sealing insert into the module cover until the locking lugs engage. 7. Check that the sealing insert is fitted correctly in the module cover. 00733286.eps Dräger X-dock 5300/6300/6600...
Dimensions (H x W x D): Order No. Name and description Master approx. 3.54 x 5.70 x 9.84 inches Dräger X-dock 5300 X-am 125 83 21 880 (120 x 130 x 250 mm) Dräger X-dock 5300 Pac 83 21 881 Module approx.
Database Management System Data logger Favourite HORN Horn Light-emitting diode Lower Explosive Limit Mass Storage Device Master SPAN Span calibration Software Response time test Time-weighted average UNDEF Unknown Unknown info Vibration ZCHECK Zero-point check ZERO Zero-point calibration Dräger X-dock 5300/6300/6600...