Manage Tests - Dräger X-dock 5300 Technical Manual

Hide thumbs Also See for X-dock 5300:
Table of Contents

Advertisement

Menu
6.2

Manage tests

In this menu, the existing tests can be edited, new tests can be
created and the standard test can be defined.
Select
> Define test.
The test overview will be displayed.
For more information see chapter 4.9 on page 14.
6.3
Test gas configuration of the station
The individual gas connections can be configured in this menu.
Select
> Gas configuration.
An overview of the test gas connections is displayed.
For more information see chapter 4.7 on page 10.
6.4
System configuration
6.4.1
Set Date & time
NOTICE
i
i
If the station is connected to a server, only place and
format can be changed. The station synchronises time
and date automatically.
The following settings can be made in this menu:
Time zone
Date format
Date
Time
Location
To set the date and time:
1. Select
> System configuration > Date & time.
2. Make the desired settings.
3. Confirm settings with OK.
6.4.2
Setting the language
1. Select
> System configuration > Language.
2. Select desired language.
3. Confirm selection with OK.
6.4.3
Network
NOTICE
i
i
If the IP or DBMS address is changed, the station must
be re-started to make the change active.
DNS is supported in the network configuration.
To set System address:
1. Select
> System configuration > Network > System
address.
2. Select dynamic or static IP address.
3. Where necessary, enter the IP of the system address.
4. Where necessary, enter the Subnet Mask (default:
255.255.255.0).
20
To set the central database address:
1. Select
> System configuration > Network > DBMS
address.
2. Enter the server IP.
6.4.4
Cleaning the database
1. Select
> System configuration > Database > Clean
up.
2. Select Start clean up.
3. Confirm deletion of the database entries with OK.
The progress of the deletion is displayed.
NOTICE
i
i
The database entries, data loggers and certificates are
deleted.
6.4.5
Display info screen
1. Select
> System configuration > Info screen.
The following information is displayed:
Application version
Hardware version
Kernel version
MAC address
IP address
Part number
Serial number
6.4.6
Performing an update
To perform a firmware update, see chapter 9.4 on page 26.
6.4.7
Service interval
1. Select
> System configuration > Service interval.
2. This function displays how many days have passed since
the last service was performed. If the service interval has
been exceeded, an additional check box to deactivate the
note for the self check for 365 days is displayed.
6.4.8
Station options
1. Select
> System configuration > Station options.
2. Activate or deactivate desired options:
Direct charge
Activate this option so that the gas monitors are not
switched on after insertion.
In the case of X-am 125+ modules, the gas monitors
are charged directly.
Enable vibration test
Activate this option to test the vibration function of the
gas monitors.
Enable autom. logout
Activate this option to log a user out automatically once
a test has been carried out.