Accuracy Check; Set Mode - Dräger Alcotest 6510 Instructions For Use Manual

Breath alcohol screening device
Hide thumbs Also See for Alcotest 6510:
Table of Contents

Advertisement

Accuracy check

The calibration parameters are not changed with this procedure!
Enter the "Accuracy check" item by using the
gas to be used.
Enter "Next" item by using the
When "Supply gas" is shown supply respective gas.
A minimum flow rate of 3 L/min obtain "Gas flow ok" on display.
The instrument performs the check automatically.
The result is displayed together with the check value.

Set mode

Enter "Set mode" item by using the
Change by pressing
or
M
XXXX: according to local settings
Back by enter the "Back to menu" item by using the
18
- button. The display shows the calibration
OK
- button.
OK
- button.
OK
and choice with
- button.
OK
- button.
OK
Accuracy check
Back to menu
Check value:
35 µg/100 ml
Gas Type:
Dry gas
Next
Check value : 47.6
Result :
47.0
µg/100mL
Unit :
Set mode
Back to menu
XXXX Traffic
XXXX Airline