Maintenance
Overview of charge status LEDs
Colour
Status
green
on continuously
green
flashing
red
flashing
9.4

Perform a firmware update

CAUTION
!
The station power supply must not be disconnected
during the installation process. The station may be
damaged if this is not observed.
NOTICE
i
i
The station does not support any USB data storage
device with an NTFS file system.
1. Download the firmware update from the internet:
a. Go to www.draeger.com.
b. Go to the X-dock product page and unzip the firmware
update to the root directory of an empty USB data
storage device.
CAUTION
!
The USB data storage device must not contain older
firmware files!
2. Connect USB data storage device with the firmware update
to the USB port of the station.
The USB icon is displayed in the status bar.
3. Select
>
configuration > Update.
A list of all firmware updates
available on the USB storage
device will be displayed.
4. Select
the
desired
update from the list. The selected
firmware update is marked blue.
5. Start the firmware update using OK.
26
Meaning
Charge status 100 %
Battery charging.
Charging error
System
firmware
The installation progress will be
displayed:
6. Following successful transfer to the station, the start is re-
started automatically and the firmware update is installed
immediately afterwards. During the installation process,
the status LEDs on the modules will be white.
7. After the installation is complete, the station changes to
operating mode. The station is ready for operation.