Siemens milltronics BW500 Instruction Manual page 50

Hide thumbs Also See for milltronics BW500:
Table of Contents

Advertisement

Connections
If the span selection is to be done by remote contact, the following connections would
apply. Otherwise, no additional connections to the BW500 are required.
Multispan Selection of Spans 1 and 2
*Remote contact can be from relay or open collector switch.
Programming
Access P365 and enter EDIT mode
P365 Multispan
Select [1-8]
Span 1 will have already been set as part of the Start Up and initial calibration. Therefore,
select 2.
Access P017 and enter EDIT mode
P017 Test Load: Weight MS2
Enter test load
Enter the test load value, and press
To do a span calibration for another condition, (i.e. span 3 or 4 etc.), access P365 and
repeat these steps for each condition. As with any initial span, follow the span calibration
for each multispan with a material test and factoring.
To use remote span selection, auxilliary Inputs, 1 and/or 2 or 3, are programmed to read
the contact state as the span selection. Remote selection overrides the keypad (or
Dolphin Plus) selection. The auxilliary inputs override the keypad selection.
Page 44
Milltronics BW500 – INSTRUCTION MANUAL
Multispan Selection of Spans 1 to 8
*
E
0
E
0
to do a span calibration.
7ML19985DK03

Advertisement

Table of Contents
loading