This is the end of zero calibration. Proceed with zero or span recalibration or
Note:
return to RUN.

Initial Zero

Perform an initial zero if necessary when a calibration is out of range message is shown.
Access P377 and enter EDIT mode
P377 Initial Zero
Enter 1 to start initial Zero
Press
Zero Calibration. Current Zero
Clear belt. Press ENTER to start
Press
Initial Zero Calibration in progress
Current Reading:
Calibration complete. Deviation
Press ENTER to accept value
Press
Zero Calibration. Current Zero
Clear belt. Press ENTER to start
This is the end of zero calibration. Proceed with span recalibration or return to
Note:
RUN.

Direct Zero

Use direct zero entry (P367) when replacing software or hardware, if it is not convenient
to perform an initial zero. A record of the last valid zero count is required.
Access P367 and enter EDIT mode
P367 Direct Zero Entry
Enter Zero Count
7ML19985DK03
Milltronics BW500 – INSTRUCTION MANUAL
E
0
530560
e.g. the current zero
the zero count being calculated while
#####
calibration is in progress
e.g. the deviation from the previous
0.00
zero
551413
e.g. the new zero count if accepted
if unacceptable, press
551413
e.g. the current zero count
E
0
to restart
Page 39