P634
Communication Totalizer Resolution
Used to set the number of fixed decimal places for Total 1 and Total 2 for SmartLinx or
Modbus communication.
Entry:
P634 Index
Primary Index 1
Primary Index 2
With 3 decimal places set, the largest readable value is 2, 147, 483.638. With 2 decimal
places set, the largest readable value is 21, 474, 836.38. With 1 or 0 decimal places set, the
largest readable value is 100, 000, 000.
This parameter is only relevant if viewing the totalizer value using remote
Note:
communications, such as SmartLinx or Modbus.
P635
Verification Totalizer
Enables a dedicated internal totalizer that totals the amount of material conveyed during
a zero or span verification. It is used to verify the accuracy of the scale.
If a printer is connected to a port and the required programming is in order, a printout of
the activity is automatically done on completion of the verification process.
Page 124
Description
Total 1 for SmartLinx communication
Total 2 for SmartLinx communication