P368
Direct Span Entry
Directly enters the span reference count for the span selected, span 1 to 8 (P368-01 to -
08).
Direct entry is intended for use when replacing software or hardware and it is not
convenient to perform an initial span at that time. (
Recalibration
Refer to
P370
Zero Limit Deviation %
Sets the zero calibration deviation limit (±) from the last initial zero. If the accumulated
deviation of successive zero calibrations exceeds the limit, the zero calibration is aborted.
f
(
= 12.5)
If the legal for trade certification switch has been set, the zero limit is
Note:
±2%.
Enter the maximum allowable % deviation.
P377
Initial Zero
Resets the initial zero.
The initial zero is the reference zero to which all subsequent operator initiated zero
calibrations are compared in determining whether they have deviated beyond the zero
f
limit (P370). (
= 1)
Refer to
Note:
P388
Initial Span
Resets the initial span for the span selected, multispan 1 to 8 (P388-01 to –08).
The initial span is the reference to which all subsequent span calibrations are compared
in determining whether they have deviated beyond an accumulated ±12.5% of the initial
f
span. (
= 1)
Refer to
Note:

Linearization Parameters (P390 - P392)

These parameters are used to compensate for non-linear response of the weighing
system to the BW500. Refer to
the use of these parameters.
In the case of multispan operation, the linearizer is applied to all
Note:
spans.
7ML19985DK03
Milltronics BW500 - INSTRUCTION MANUAL
on page 34 for execution.
Initial Zero
on page 39 for execution.
Initial Span
on page 42 for execution.
Linearization
f
= 0)
on page 51 for execution, and example on
Page 117