P948
Error Log
Displays a log of the last 25 error or alarm events (P948 – 01 to – 25) that have occurred.
Event 01 is the current error.
Display:
0 = no error
# = error code; refer to
P949
Diagnostic Error Testing
Enables or disables diagnostic error checking for memory, load cell and zero speed.
f
(
= 0)
Entry:
0 = disable
1 = enable
Troubleshooting
Refer to
P950
Zero Register
Registers the number of zero calibrations that have been done since the last master
f
reset. (
= 0)
P951
Span Register
Registers the number of span calibrations for the span selected, span 1 to 8 (P951 - 01 to -
08), that have been done since the last master reset. (
P952
Design Load
Displays the value of the design load, which corresponds to the full scale value for alarm
and mA output functions. The design load is calculated, based on the design rate and
f
design speed. (
= 0.00)
P999
Master Reset
Resets parameters and totalizers to their factory setting. (
Enter 9 to execute the reset.