Maintenance
3.
Press it downwards until
it is in position. Be careful not
to break the holder at the front
of the Bottle holder (marked by
being encircled in Fig. 2).
Replace broken Tube Holders on the
Sample Carousel
1.
Unscrew the center screw.
Fig. 3.
2.
Lift the upper part of the
Sample Carousel. Fig. 4.
3.
Remove the broken Tube
holder and replace it with a
new one. Fig. 5.
4.
Remove the Tube holder
guide from inside the Sample
Carousel bowl. Fig. 6.
P a g e 1 1 . 2 2
Phadia 100 - User Manual
Fig.2
Fig.3
Fig.4
Fig.5
Fig.6
12-3501-20/12