Mounting
Wetted materials
Diaphragm protec-
tion
Installation location
Moisture
10
4 Mounting
4.1 General instructions
Check if the wetted materials such as seal, process fitting etc.
are suitable for the existing process conditions, such as
product, temperature etc.
Caution:
After removing the diaphragm protection, the diaphragm must
not be impressed.
Select an installation position you can easily reach for
mounting and connecting as well as later retrofitting of an
indicating and adjustment module. The housing can be rotated
by 330° without the use of any tools. You can also install the
indicating and adjustment module in four different positions
(each displaced by 90°).
Use the recommended cable (see chapter "Connecting to
power supply") and tighten the cable gland.
You can give your VEGABAR 53 additional protection against
moisture penetration by leading the connection cable down-
ward in front of the cable entry. Rain and condensation water
can thus drain off. This applies mainly to mounting outdoors, in
areas where moisture is expected (e.g. by cleaning processes)
or on cooled or heated vessels.
Fig. 2: Measures against moisture penetration
VEGABAR 53 - Profibus PA