Version for oxygen applications
- without/with PLICSCOM
Process conditions
The specifications of the pressure stage are used as an overview. The specifications on the
type plate are applicable.
Pressure stage, process fitting
- Thread
- Tri-Clamp 1 1/2" 316L
- Tri-Clamp 2 " 316L
- Hygienic fitting LA 316L
Product temperature depending on seal and O-ring
- Viton
- EPDM
- NBR
Product temperature with connection G1A front flush acc. to EHEDG
- Viton, EPDM, NBR
Product temperature hygienic fittings
- Tri-Clamp
- hygienic fitting LA, LD
Vibration resistance
44
13)
-40 ... +60°C (-40 ... +140°F)
see type label
PN 16
PN 10
PN 40
-20 ... +105°C (-4 ... +221°F)
-40 ... +105°C (-40 ... +221°F)
-25 ... +105°C (-13 ... +221°F)
-10 ... +150°C (+14 ... +302°F)
-20 ... +150°C (-4 ... +302°F)
-20 ... +120°C (-4 ... 248°F)
mechanical vibrations with 4 g and 5 ... 100 Hz
up to +60°C (+140°F).
13)
Version for oxygen applications up to +60°C (+140°F).
14)
Tested acc. to the regulations of German Lloyd, GL directive 2
15)
14)
VEGABAR 53 - Profibus PA
15)