I/O Module Self-Test - Fluke 9100A Operation Manual

9100 series
Hide thumbs Also See for 9100A:
Table of Contents

Advertisement

The I/O module self-test is a pass/fail check that the module
concerned is connected to and communicating with the system.
To execute the test:
Press the MAIN MENU key, and use the left arrow
1 .
key to move the cursor to the left-most field.
2 .
Press the SELFTEST softkey.
3 .
Move the cursor to the next field, and press the I/O
MOD softkey.
4.
Move the cursor to the next field and identify the
module to be tested by a number (1 through 4).
For example, if module 2 is to be tested, press the
"2" key (a white key on the operator's keypad). The
display should read:
MAIN:
5.
Press the ENTER key. If the self-test fails, a failure
message will appear on the operator's display. If
this happens, check the I/O module connections and
repeat the test carefully following any instructions.
If the test still fails, the I/O module requires service.
I/O MOD 2
4 . 2 . 3 .
4-5