Fluke 9100A Operation Manual page 46

9100 series
Hide thumbs Also See for 9100A:
Table of Contents

Advertisement

After each power-up or reset before you test or trouble-
shoot a UUT:
Calibrate and configure the system for the type of UUT
you are about to test or troubleshoot.
l
Ensure that self-tests have been performed on the pod,
probe, and I/O modules.
Verify that the system mode is appropriate and that your
user disk is in the correct drive.
If you reset or change any part of the system:
l
The self-tests will have to be repeated and the system
recalibrated. You can save new calibration data on the user
disk, writing over the old data.
Test and troubleshooting programs and the system config-
uration remain valid.
If you change UUTs from one type to another:
l
You may have to change pods or I/O modules. If so,
repeat the self-tests and calibrations with the new pod.
l
System settings are probably different for the new UUT.
Restore settings from the new UUT directory to recon-
figure the system.
4-12

Advertisement

Table of Contents
loading

This manual is also suitable for:

9105a

Table of Contents