SUMMARY OF GFI KEY OPERATIONS

To run GFI from a specified location:
Disable the UUT microprocessor.
l
Connect the pod to the UUT.
Press the GFI key and use the left arrow key to move the
cursor to the left-most field.
l
Press the RUN softkey.
l
Type the UUT directory name at the second field.
Type the component name at the third field.
Type the component pin name at the last field.
l
Press the ENTER key.
To display the GFI summary:
Press the GFI key and use the left arrow key to move the
cursor to the left-most field.
l
Press the SUMMARY softkey.
l
Press the ENTER key.
To see the GFI suggestion list:
l
Press the GFI key and use the left arrow key to move the
cursor to the left-most field.
l
Press the SUGGEST softkey.
l
Press the ENTER key.