Fluke Hart Scientific 914 Series User Manual
Fluke Hart Scientific 914 Series User Manual

Fluke Hart Scientific 914 Series User Manual

Field metrology well

Advertisement

Quick Links

914X Series
Field Metrology Well
User's Guide
Revision 7A1901-EN
MyFlukeStore
1.888.610.7664
www.
.com
Shop for Fluke products online at:

Advertisement

Table of Contents
loading

Summary of Contents for Fluke Hart Scientific 914 Series

  • Page 1 914X Series Field Metrology Well User’s Guide Revision 7A1901-EN MyFlukeStore 1.888.610.7664 www. .com Shop for Fluke products online at:...
  • Page 2 Limited Warranty & Limitation of Liability Each product from Fluke Corporation, Hart Scientific Division (“Hart”) is warranted to be free from defects in material and workmanship under normal use and service. The warranty period is one year for Field Metrology Wells. The war- ranty period begins on the date of the shipment.
  • Page 3: Table Of Contents

    Menu Structure ...............29 Temp Setup Menu ................... 29 Prog Menu ....................30 4.2.1 Switch Test Parameters .................31 4.2.2 Switch Test Description .................31 System Menu ..................33 Input Setup (-P only) ................34 MyFlukeStore 1.888.610.7664 www. .com Shop for Fluke products online at:...
  • Page 4 Maintenance ................35 Field Metrology Well Performance Analysis ........... 35 MyFlukeStore 1.888.610.7664 www. .com Shop for Fluke products online at:...
  • Page 5 Tables Table 1 Symbols used ..................3 Table 2 Base Unit Specifications ..............13 Table 3 -P Option Specifications ..............14 MyFlukeStore 1.888.610.7664 www. .com Shop for Fluke products online at:...
  • Page 6 Figure 12 Main Menu - Prog Menu ..............30 Figure 13 Auto and manual switch test operation example ......32 Figure 14 Main Menu - System Menu .............. 33 Figure 15 Main Menu - Input Setup ..............34 MyFlukeStore 1.888.610.7664 www. .com Shop for Fluke products online at:...
  • Page 7: Before You Start

    Japanese, Chinese, German, Spanish, French, Russian, and Italian. The instrument’s rugged design and special features make them ideal for the field or the labora- tory. With proper use, the instrument provides continued accurate calibration of temperature MyFlukeStore 1.888.610.7664 www. .com Shop for Fluke products online at:...
  • Page 8: Unpacking

    6-pin DIN Connector (-P model only) ● Test Lead Kit (-P model only) ● Clamp-on ferrites (3) [-P model only] ● Tongs (insert removal tool) ● 9930 Interface-it Software and User’s Guide ● MyFlukeStore 1.888.610.7664 www. .com Shop for Fluke products online at:...
  • Page 9: Symbols Used

    Table 1 lists the International Electrical Symbols. Some or all of these symbols may be used on the instrument or in this guide. Table 1 Symbols used Symbol Description AC (Alternating Current) AC-DC Battery Complies with European Union directives Double Insulated MyFlukeStore 1.888.610.7664 www. .com Shop for Fluke products online at:...
  • Page 10: Safety Information

    The following definitions apply to the terms “Warning” and “Caution”. “Warning” identifies conditions and actions that may pose hazards to the user. ● “Caution” identifies conditions and actions that may damage the instrument being used. ● MyFlukeStore 1.888.610.7664 www. .com Shop for Fluke products online at:...
  • Page 11: Warnings

    Do not use the instrument if it operates abnormally. Protection may be impaired. When in doubt, have the instrument serviced. Do not apply more than the rated voltage, as marked on the instrument, between terminals or between any terminal and earth ground. MyFlukeStore 1.888.610.7664 www. .com Shop for Fluke products online at:...
  • Page 12 The receptacle must be installed in accordance with local codes and ordinances. Consult a qualified electrician. DO NOT use an extension cord or adapter plug. MyFlukeStore 1.888.610.7664 www. .com Shop for Fluke products online at:...
  • Page 13: Cautions

    Unless recalibrating the instrument DO NOT change the values of the calibration constants from the factory set values. The correct setting of these parameters is important to the safety and proper operation of the calibrator. MyFlukeStore 1.888.610.7664 www. .com Shop for Fluke products online at:...
  • Page 14: Ce Comments

    (EM) immunity in environments of excessive EM interference. During EMC testing we found that ferrites clamped around probe cables for the Reference PRT, the PRT/RTD input, and the thermocouple (TC) input reduced the risk the EM interference affects measurements. Therefore, MyFlukeStore 1.888.610.7664 www. .com Shop for Fluke products online at:...
  • Page 15: Emission Testing

    EN 61010-1 and the EN 61010-2-010 standards. Authorized Service Centers Please contact one of the following Authorized Service Centers to coordinate service on your Hart product: MyFlukeStore 1.888.610.7664 www. .com Shop for Fluke products online at:...
  • Page 16 Authorized Service Centers When contacting a Service Centers for support, please have the following information available: Model Number ● Serial Number ● Voltage ● Complete description of the problem ● MyFlukeStore 1.888.610.7664 www. .com Shop for Fluke products online at:...
  • Page 17 MyFlukeStore 1.888.610.7664 www. .com Shop for Fluke products online at:...
  • Page 18 MyFlukeStore 1.888.610.7664 www. .com Shop for Fluke products online at:...
  • Page 19: Specifications And Environmental Conditions

    LCD, °C or °F user-selectable Key Pad Arrows, Menu, Enter, Exit, 4 soft keys Size (H x W x D) 290 mm x 185 mm x 295 mm (11.4 x 7.3 x 11.6 in) MyFlukeStore 1.888.610.7664 www. .com Shop for Fluke products online at:...
  • Page 20 0 ohms to 25 ohms: ±0.002 ohms ‡ 25 ohms to 400 ohms: ±80 ppm of reading RTD Characterizations PT-100 (385),(JIS),(3926), NI-120, Resistance RTD Measurement Capability 2-,3-,4-wire RTD w\ Jumpers only RTD Connection 4 terminal input MyFlukeStore 1.888.610.7664 www. .com Shop for Fluke products online at:...
  • Page 21: Environmental Conditions

    0 % to 90 % (non-condensing) ● mains voltage: within ±10% of nominal ● vibrations in the calibration environment should be minimized ● altitude: less than 2,000 meters ● indoor use only ● MyFlukeStore 1.888.610.7664 www. .com Shop for Fluke products online at:...
  • Page 22 MyFlukeStore 1.888.610.7664 www. .com Shop for Fluke products online at:...
  • Page 23: Quick Start

    TER” to save the desired set-point and enable the instrument. After five (5) seconds, the instru- ment should start to operate normally and heat or cool to the designated set-point. MyFlukeStore 1.888.610.7664 www. .com Shop for Fluke products online at:...
  • Page 24: Parts And Controls

    Parts and Controls This section describes the exterior features of the Field Metrology Well. All interface and power connections are found on the front of the instrument (see Figure 2). MyFlukeStore 1.888.610.7664 www. .com Shop for Fluke products online at:...
  • Page 25: Display Panel

    The function of the keys may change depending on the menu or function that is selected. Switch Connector (9) The switch hold connector posts are located on the left side of the display panel. MyFlukeStore 1.888.610.7664 www. .com Shop for Fluke products online at:...
  • Page 26: Display

    The current set-point temperature is displayed just below the Process Temperature. Reference Thermometer Temperature (3) [-P models only] When installed, the most recent reference thermometer measurement is shown on the screen. MyFlukeStore 1.888.610.7664 www. .com Shop for Fluke products online at:...
  • Page 27: Figure 4 914X Display

    While setting up and operating the instrument, you are often required to enter or select param- eters. Editing windows appear on the screen when necessary to show the values of parameters and allow edits. Figure 4 914X display MyFlukeStore 1.888.610.7664 www. .com Shop for Fluke products online at:...
  • Page 28: Power Panel

    For the 9143 and 9144, the fuses are separate from the power connector (Figure 6 on opposite page). If necessary, fuses must be replaced according to Specifications (see Section 2.1 Specifications on page 13. MyFlukeStore 1.888.610.7664 www. .com Shop for Fluke products online at:...
  • Page 29: Figure 5 9142 Power Panel

    Quick Start Parts and Controls Figure 5 9142 power panel Figure 6 9143 and 9144 power panel MyFlukeStore 1.888.610.7664 www. .com Shop for Fluke products online at:...
  • Page 30: P Option Panel (-P Models Only)

    2 and 4 sense the potential). If a shield wire is present, it should be connected to pin 3, which is also used for the memory circuit. Pin 6 is only used for the memory circuit. MyFlukeStore 1.888.610.7664 www. .com Shop for Fluke products online at:...
  • Page 31: Figure 8 Probe Connector Wiring

    RTD is shown on the instrument. Figure 9 shows the correct wiring for a 2 or 3-wire PRT/RTD (see Section 1.5.2 Immunity Testing on page 8 for information on using clamp-on ferrites). MyFlukeStore 1.888.610.7664 www. .com Shop for Fluke products online at:...
  • Page 32: Languages

    ● Russian: Russian, English ● Asian: English, Chinese, Japanese ● 3.3.1 Language Selection Select the language to be displayed by following the steps shown in Figure 10 on opposite page. MyFlukeStore 1.888.610.7664 www. .com Shop for Fluke products online at:...
  • Page 33: Reset To English Language

    Note: The F1 and F4 English shortcut override is temporary. If you toggle the power off, the instrument jwill return to the language selected in the DISPLAY SETUP menu rather than coming up in English. MyFlukeStore 1.888.610.7664 www. .com Shop for Fluke products online at:...
  • Page 34 MyFlukeStore 1.888.610.7664 www. .com Shop for Fluke products online at:...
  • Page 35: Menu Structure

    2 Beep – Invalid key entry Code Update Mode Keys ENTER & EXIT Keys (hold during power up) <initiate code update mode> Allows instrument software to be updated Figure 11 Main Menu - Temp SetUp MyFlukeStore 1.888.610.7664 www. .com Shop for Fluke products online at:...
  • Page 36: Prog Menu

    PRINT TESTS: < NO, YES> <select> Print all test data to RS-232 port ERASE TESTS ERASE TESTS: <NO, YES> <select> Erases all test data Figure 12 Main Menu - Prog Menu MyFlukeStore 1.888.610.7664 www. .com Shop for Fluke products online at:...
  • Page 37: Switch Test Parameters

    TEMP, APPROACH LIMIT, and NO. CYCLES parameters. Set the Test Method to MANUAL. Exit to the Run Prog menu. Ensure that the Run Test is set to SWITCH TEST. Set Test Status to MyFlukeStore 1.888.610.7664 www. .com Shop for Fluke products online at:...
  • Page 38: Figure 13 Auto And Manual Switch Test Operation Example

    BAND are displayed for the user to record. The values may also be recorded inter- nally in the instrument by selecting the option to record the data (-P model only). Figure 13 Auto and manual switch test operation example MyFlukeStore 1.888.610.7664 www. .com Shop for Fluke products online at:...
  • Page 39: System Menu

    MODEL: <view only> SERIAL number: <view only> FirmWare VERsion: <view only> CALibration DATE: <view only> INPUT CALibration DATE: : <view only> (-P only) Figure 14 Main Menu - System Menu MyFlukeStore 1.888.610.7664 www. .com Shop for Fluke products online at:...
  • Page 40: Input Setup (-P Only)

    ALPHA BETA DELTA PROGRAM probe <YES, NO> <password required to program probe> TEST reference CALCulation RESISTANCE: <edit> TEMPERATURE: <calculated value displayed> <view only> Figure 15 Main Menu - Input Setup MyFlukeStore 1.888.610.7664 www. .com Shop for Fluke products online at:...
  • Page 41 Before using any cleaning or decontamination method, other than those recommended by ● Fluke’s Hart Scientific Division, users should check with an Authorized Service Center to insure the proposed method will not damage the equipment. If the instrument is used in a manner not in accordance with the equipment design, the ●...
  • Page 42 Field Metrology Well Calibration section of the Field Metrology Well Technical Guide and recalibrate the Field Me- trology Well. MyFlukeStore 1.888.610.7664 www. .com Shop for Fluke products online at:...

Table of Contents