SELF-TESTS (MAIN MENU KEY)

Three self-tests should be performed after you power-up the
system or change the I/O module, pod, or probe. Before
running the tests, check that the devices are connected correctly.
Probe
Self-Test
The probe self-test is a pass/fail check that the probe is
connected to and communicating with the system. To execute
this test:
1 .
Press the MAIN MENU key, and use the left arrow
key to move the cursor to the left-most field.
2.
Press the SELFTEST softkey.
3 .
Move the cursor to the next field, and press the
PROBE softkey. The display should read:
MAIN:
4.
Press the ENTER key to perform the self-test. If the
self-test fails, a failure message will appear on the
operator's display. If this happens, check the probe
connection and repeat the test, carefully following
any instructions. If the test still fails, the probe
requires service.