Fluke 9142 User Manual

Fluke 9142 User Manual

Field metrology well
Hide thumbs Also See for 9142:

Advertisement

9142, 9143, 9144
Field Metrology Well
User's Guide
914X____ugeng0001
Revision 991702

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the 9142 and is the answer not in the manual?

Questions and answers

Summary of Contents for Fluke 9142

  • Page 1 9142, 9143, 9144 Field Metrology Well User’s Guide 914X____ugeng0001 Revision 991702...
  • Page 2 Limited Warranty & Limitation of Liability Each product from Fluke Corporation, Hart Scientific Division (“Hart”) is warranted to be free from defects in material and workmanship under normal use and service. The warranty period is one year for Field Me- trology Wells.
  • Page 3: Table Of Contents

    Table of Contents Before You Start ...............1 1.1 Introduction ....................1 1.2 Unpacking ....................2 1.3 Symbols Used ................... 3 1.4 Safety Information ..................4 1.4.1 Warnings ......................5 1.4.2 Cautions ......................7 1.5 CE Comments ................... 8 1.5.1 EMC Directive ....................8 1.5.2 Immunity Testing .....................8 1.5.3 Emission Testing ....................9 1.5.4 Low Voltage Directive (Safety) ................9 1.6 Authorized Service Centers ..............
  • Page 4 Maintenance ................33 5.1 Field Metrology Well Performance Analysis ........... 33...
  • Page 5 Tables Table 1 Symbols used ..................3 Table 2 Base Unit Specifications ..............11 Table 3 -P Option Specifications ..............12...
  • Page 6 Figures Figure 1 Clamp-on ferrite installation ..............9 Figure 2 914X Field Metrology Well ..............16 Figure 3 Display panel and keys ..............18 Figure 4 914X display ..................19 Figure 5 9142 power panel ................21 Figure 6 9143 and 9144 power panel .............. 21 Figure 7 -P option panel ................. 22 Figure 8 Probe connector wiring ..............23 Figure 9 Jumper locations for 3-wire and 2-wire connections ......24 Figure 10 Steps to language selection ............25 Figure 11 Main Menu - Temp SetUp ..............27 Figure 12 Main Menu - Prog Menu ..............28 Figure 13 Auto and manual switch test operation example ......30 Figure 14 Main Menu - System Menu .............. 31 Figure 15 Main Menu - Input Setup ..............32...
  • Page 7: Before You Start

    Before You Start Introduction Field Metrology Wells (9142, 9143, and 9144) are designed to be reliable, stable heat sources that can be used in the field or laboratory. They offer accuracy, portability, and speed for nearly every field calibration application. The instruments have been designed with the field user in mind and are easy to use while maintaining stability, uniformity, and accuracy comparable to some laboratory instruments.
  • Page 8: Unpacking

    If there is shipping damage, notify the carrier immediately. Verify that the following components are present: 9142 9142 Field Metrology Well ● ● 9142-INSX Insert (X=A, B, C, D, E, or F) ● ● Power Cord ● ● RS-232 Cable ●...
  • Page 9: Symbols Used

    Before You Start Symbols Used 9144 9144 Field Metrology Well ● ● 9144-INSX Insert (X=A, B, C, D, E, or F) ● ● Power Cord ● ● RS-232 Cable ● ● User Guide ● ● Technical Manual CD ● ● Report of Calibration and calibration label ●...
  • Page 10: Safety Information

    914X Field Metrology Wells Safety Information Symbol Description Electric Shock Fuse PE Ground Hot Surface (Burn Hazard) Read the User’s Guide (Important Information) Canadian Standards Association C-TICK Australian EMC mark The European Waste Electrical and Electronic Equipment (WEEE) Directive (2002/96/ EC) mark.
  • Page 11: Warnings

    Before You Start Safety Information 1.4.1 Warnings To avoid personal injury, follow these guidelines. GENERAL DO NOT use this instrument in environments other than those listed in the User’s Guide. Inspect the instrument for damage before each use. Inspect the case. Look for cracks or missing plastic.
  • Page 12 914X Field Metrology Wells Safety Information Never touch the probes to a voltage source when the test leads are plugged into the current terminals. Select the proper function and range for each measurement. Disconnect the test leads before changing to another measure or source function. DO NOT operate the Field Metrology Well around explosive gas, vapor, or dust.
  • Page 13: Cautions

    Before You Start Safety Information If supplied with user accessible fuses, always replace the fuse with one of the same rating, volt- age, and type. Always replace the power cord with an approved cord of the correct rating and type. HIGH VOLTAGE is used in the operation of this equipment.
  • Page 14: Ce Comments

    914X Field Metrology Wells CE Comments DO NOT allow the probe sheath or inserts to drop into the well. This type of action can cause a shock to the sensor and affect the calibration. The instrument and any thermometer probes used with it are sensitive instruments that can be easily damaged.
  • Page 15: Emission Testing

    EN 61010-1 and the EN 61010-2-010 standards. Authorized Service Centers Please contact one of the following Authorized Service Centers to coordinate service on your Hart product: Fluke Corporation Hart Scientific Division 799 E. Utah Valley Drive American Fork, UT 84003-9775 Phone: +1.801.763.1600...
  • Page 16 22 Jianguomenwai Dajie Chao Yang District Beijing 100004, PRC CHINA Phone: +86-10-6-512-3436 Telefax: +86-10-6-512-3437 E-mail: xingye.han@fluke.com.cn Fluke South East Asia Pte Ltd. Fluke ASEAN Regional Office Service Center 60 Alexandra Terrace #03-16 The Comtech (Lobby D) 118502 SINGAPORE Phone: +65-6799-5588 Telefax: +65-6799-5589 E-mail: anthony.ng@fluke.com...
  • Page 17: Specifications And Environmental Conditions

    Specifications and Environmental Conditions Specifications Specifications and Environmental Conditions Specifications Table 2 Base Unit Specifications Base Unit Specifications 9142 9143 9144 Temperature Range at –25 °C to 150 °C 33 °C to 350 °C 50 °C to 660 °C 23 °C (-13 °F to 302 °F) (91 °F to 662 °F) (122 °F to 1220 °F)
  • Page 18 914X Field Metrology Wells Specifications Base Unit Specifications 9142 9143 9144 Weight 8.16 kg (18 lbs) 7.3 kg (16 lbs)  7.7 kg (17 lbs)  Power Requirements 100 V to 115 V (± 10 %) 100 V to 115 V (± 10 %), 50/60 Hz, 1400 W 50/60 Hz, 575 W 230 V (± 10%), 50/60 Hz, 1800 W...
  • Page 19: Environmental Conditions

    Specifications and Environmental Conditions Environmental Conditions -P Specifications Built-in TC Thermometer Readout Accuracy Type J: ± 0.7 °C at 660 °C Type K: ± 0.8 °C at 660 °C Type T: ± 0.8 °C at 400 °C Type E: ± 0.7 °C at 660 °C Type R: ±...
  • Page 21: Quick Start

    Quick Start Setup Quick Start Setup Note: The instrument will not heat, cool, or control until the “SET PT.” parameter is “Enabled”. Place the calibrator on a flat surface with at least 6 inches of free space around the instrument. Overhead clearance is required.
  • Page 22: Parts And Controls

    914X Field Metrology Wells Parts and Controls Figure 2 914X Field Metrology Well Parts and Controls This section describes the exterior features of the Field Metrology Well. All interface and power connections are found on the front of the instrument (see Figure 2).
  • Page 23: Display Panel

    Quick Start Parts and Controls 3.2.1 Display Panel Figure 3 on next page shows the layout of the display panel. Display (1) The display is a 240 x 160 pixel monochrome graphics LCD device with a bright LED back- light. The display is used to show current control temperature, measurements, status informa- tion, operating parameters, and soft key functions.
  • Page 24: Display

    914X Field Metrology Wells Parts and Controls Block Temperature Indicator (10) [US Patent 7,561,058] The Block Temperature Indicator lamp allows users to know when the block temperature is safe (50°C to 60°C) to remove inserts or move the Field Metrology Well. The indicator light is lit continuously once the block has exceeded approximately 50°C (varies 50°C to 60°C).
  • Page 25: Figure 4 914X Display

    Quick Start Parts and Controls Stability Status (4) On the right hand side of the screen, you will find a graph displaying the current status of the stability of the Field Metrology Well. Heating/Cooling Status (5) Just below the stability graph there is a bar graph that will indicate HEATING, COOLING, or CUTOUT.
  • Page 26: Power Panel

    Power Switch (2) For the 9142, the power switch is located on the power entry module of the unit at the lower center of the power panel. For the 9143 and 9144, the power switch is located between the RS-232 and the fuses.
  • Page 27: Figure 5 9142 Power Panel

    Quick Start Parts and Controls Figure 5 9142 power panel Figure 6 9143 and 9144 power panel...
  • Page 28: P Option Panel (-P Models Only)

    914X Field Metrology Wells Parts and Controls 3.2.4 -P Option Panel (-P models only) The –P (process version) panel is the readout portion of the instrument and is only available with –P models. MEASURE mA LOOP REF PRT 50 mA F 250 V 30 V, 24 mA MAX ALL TERMINALS Figure 7...
  • Page 29: Figure 8 Probe Connector Wiring

    Quick Start Parts and Controls Probe Connector Shield RTD Sensor Figure 8 Probe connector wiring A two-wire probe can also be used with the reference thermometer. It is connected by attach- ing one wire to both pins 1 and 2 of the plug and the other wire to both pins 4 and 5. If a shield wire is present, it should be connected to pin 3.
  • Page 30: Languages

    914X Field Metrology Wells Languages Figure 9 Jumper locations for 3-wire and 2-wire connections Thermocouple (TC) Connector (4) The TC connector allows for the use of subminiature TC connectors (see CE Comments on page 8 for information on using clamp-on ferrites). Fuse (5) Fuse for the 4-20 mA circuit.
  • Page 31: Reset To English Language

    Quick Start Languages Figure 10 Steps to language selection 3.3.2 Reset to English Language If you are in a language and need a short cut exit, press F1 and F4 simultaneously to reset the display to English. To reset to your originally selected language after resetting to English, follow the steps in Figure 10 on this page.
  • Page 33: Menu Structure

    Menu Structure Temp Setup Menu Menu Structure Temp Setup Menu 9142/9143/9144 MAIN MENU TEMP SETUP PROG MENU SYSTEM MENU INPUT SETUP (-P only) SETUP SCAN RATE: <edit> Degrees per minute at which the Field Metrology Well increases or decreases temperature STABLE LIMIT: <edit>...
  • Page 34: Prog Menu

    914X Field Metrology Wells Prog Menu Prog Menu 9142/9143/9144 MAIN MENU TEMP SETUP PROG MENU SYSTEM MENU INPUT SETUP (-P only) RUN PROGram TEST STATUS: <OFF, RUN> <select> Changes status of current test RUN TEST: <RAMP/SOAK, SWITCH TEST> <select> Runs selected test RECORD DATA: <YES, NO>...
  • Page 35: Switch Test Parameters

    Menu Structure Prog Menu 4.2.1 Switch Test Parameters SWITCH TEMP The SWITCH TEMP parameter is the nominal change temperature of the switch. UPPER TEMP The UPPER TEMP parameter is the temperature during a cycle at which the Field Metrology Well begins to heat or cool at the rate specified in “Scan Rate” found in MAIN MENU|TEMP SETUP|SETUP|SCAN RATE.
  • Page 36: Figure 13 Auto And Manual Switch Test Operation Example

    914X Field Metrology Wells Prog Menu RUN. Press Enter and the instrument will engage and start the test within a few seconds. Exit to the main screen to view the test progress, refer to the Menu Structure. When the switch resets, the test completes and the values of the switch OPEN, switch CLOSE, and switch BAND are displayed for the user to record.
  • Page 37: System Menu

    Menu Structure System Menu System Menu 9142/9143/9144 MAIN MENU TEMP SETUP PROG MENU SYSTEM MENU INPUT SETUP (-P only) SYSTEM SETUP DISPLaY SETUP LANGUAGE: <English, Japanese, Chinese, German, Spanish, French, Russian, Italian> <select> DECIMAL: <PERIOD, COMMA> <select> KEY AUDIO: <ON, OFF> <select>...
  • Page 38: Input Setup (-P Only)

    914X Field Metrology Wells Input Setup (-P only) Input Setup (-P only) 9142/9143/9144 MAIN MENU TEMP SETUP PROG MENU SYSTEM MENU INPUT SETUP (-P only) SELECT INPUT SENSOR TYPE: <RTD, TC, mA> <select> SETUP INPUT RTD SETUP WIRES: <4, 2, 3> <select>...
  • Page 39 ● ● Fluke’s Hart Scientific Division, users should check with an Authorized Service Center to insure the proposed method will not damage the equipment. If the instrument is used in a manner not in accordance with the equipment design, the ●...
  • Page 40 914X Field Metrology Wells Field Metrology Well Performance Analysis Accuracy Drift The display temperature of the instrument will drift over time. This is due to a variety of factors affecting the temperature control PRT. Any PRT is subject to changes depending on how it is used and the environment it is used in.

This manual is also suitable for:

91439144914 series

Table of Contents