Program; Storage Of Programs - Hermle Z 207 A Operating Manual

Small centrifuge
Table of Contents

Advertisement

OPERATING

2.6 Program

2.6.1 Storage of Programs

The program stores up to 99 runs, with all relevant parameters, including the used rotors. Any free
program number is available and can be retrieved.
Put the rotor into the centrifuge. By pressing the key, "prog" (11), in the display "time" (A-3) appears
the word "program", (see Figure 15). With the control field (1), choose the desired program number.
If a program number is already occupied, in the display "rpm/rcf" (A-1), the words "rotor" (M3) and
"221.57" (M4) will appear, (see Figure 15). Free program numbers will appear as 0.
Figure 15
Close the lid of the centrifuge, now proceed as described above, to set all important run parameters.
If the lid isn´t closed, when storing the program in the display "rpm/rcf" (A-1), flashes alternately the
word "FirSt" and "CLOSE Lid" (see Figure 16). When starting the run without storing the program, in
the display "rpm/rcf" (A-1), flashes alternately the word "First" and "PrESS StoreE", (see Figure 17).
Figure 16
Figure 17
For alteration of data, press the key "store" (12), for approx. 1 second. If the program is stored
correctly, the word "StorE" appears in the display "rpm/rcf" (A-1). As a result, the word "program"
(M10) disappears. As soon as the key "store" (12) is no longer displayed, the word "programm xx"
(M10) reappears, (the xx stands for the chosen program place).
If all program numbers are occupied, take an old number that is not needed any longer and replace
it with the new parameters.
Z207A_V1.17
16
_eng
© Hermle Labortechnik GmbH