OPERATING

2. OPERATING

2.1 Mounting and Loading the Rotor

2.1.1 Installation of Rotors

Clean the drive shaft and the rotor shaft with a clean, grease-free piece of cloth. Place the rotor onto
the drive shaft, (see Figure 5). Please be sure that the rotor is fully installed onto the motor shaft.
Figure 5
Hold the rotor with one hand and secure the rotor to the shaft, by turning the rotor nut clockwise (see
Figure 6).
ATTENTION:
For safety, always ensure that the rotor nut is tightened before each run!
Do not operate the centrifuge with rotors or adapters that show any signs of corrosion or
mechanical damage.
Do not operate with extremely corrosive substances, which could damage the rotor and the
centrifuge