Sartorius Signum 1 Operating Instruction page 57

Complete scales
Hide thumbs Also See for Signum 1:
Table of Contents

Advertisement

Example:
Measuring 25 m electrical cable.
Settings (changes in the factory settings required for this example):
Application parameters: Application: Neutral Measurement
Setup: Printout:
PrtProt
:
7.6.
(see "Configuration" for options)
)
248
O
; then select menu line item of your choice
1 Place empty container on the
platform
2 Tare the scale
Note: If the automatic tare
function is enabled, you do
not need to press the ) key
to tare the scale; the tare
weight is saved automatically
when you place the container
on the platform
Signum 3 only:
3 Enter the weight of 1 meter
of cable using the keypad
(in this example, 248 g)
4 Save value entered as
reference weight.
p
nRef
+
1 o
wRef
+
0.248 kg
G#
+
6.794 kg
T
+
0.541 kg
N
+
6.253 kg
Qnt
25 o
--------------------
5 Place the desired amount of
cable in the container
Read the result
6 Print the result
Configured printout:
see page 80
57

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents