Error Codes:
Pos.
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22
S t a t * * * * * E r r * * # # * * * * CR LF
or
S t a t * * * * * E r r * # # # * * * * CR LF
*:
Space
#:
Error code number (2 or 3 digits)
ID Code Characters
ID character I
Meaning
G#
Gross value
N
Net value
T
Application tare memory 1
T
Application tare memory 2
Diff
Difference from calibration value
Targ.
Exact calibration weight
Nom.
Exact calibration weight
using SBI output
nRef
Reference sample quantity
pRef
Reference percentage
wRef
Reference sample weight
Qnt
Result from Counting application
Result from Counting (piece count) and
Neutral Measurement applications
mDef
Target value for Animal weighing
x-Net
Result from Animal Weighing
Setp
Target value for Checkweighing
W.Diff
Absolute difference (e.g., in kg)
in Checkweighing
Lim
Deviation in % in Checkweighing
Max
Upper limit for Checkweighing
Min
Lower limit for Checkweighing
Stat
Status
Classx
Classification
Limx
Class limit
D
Percentage (as loss)
Prc
Percentage (as residue)
Wxx%
Reference percentage weight
Compxxx
Component xxx
Cont.T
Contents of the tare memory
in Net-total Formulation
Tot. cp
Total weight
in Net-total Formulation
PT2
Preset tare
n
Transaction counter
*G
Sum of gross weights
in Totalizing
*N
Sum of net weights
in Totalizing
Ser.no
Serial number of the platform
or display and control unit
88
Example (output of value: +1255.7 g):
Pos.
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22
G # * * * * + * * * 1 2 5 5 . 7 * g * * CR LF
Position 1–6:
ID code character, right-justified with spaces
Position 7:
Plus or minus sign or space
Position 8:
Space
Positions 9–16:
Weight value with decimal point;
leading zeros are output as spaces.
Position 17:
Space
Positions 18–20:
Unit symbol or space
Position 21:
Carriage return
Position 22:
Line feed
If the weight value is output with 10-fold increased resolution,
!
this value is not permitted to be printed or saved in a weighing
instrument operated in legal metrology in the SBI mode.
In this case, the unit symbol is not included with output.
Configuring the Data Interface
as a Printer Port ( printer )
You can connect one or two strip printers or one or two label
printers to the Signum. Configure the COM1 and UniCOM
interfaces as printer ports under the "
There are several actions that generate the command for
outputting data to the printer port:
– Pressing the p key. If the operating menu is active,
all menu settings under the active menu level are printed.
– On receipt of the SBI command
"Esc k P _". For details, see "Data Input Format" in this chapter.
– In some applications, pressing a given key (e.g., to save a value
or start a routine) also generates a print command. In this case,
a configurable printout is generated with application-specific data.
The P and b symbols are displayed when data is being output
to the printer port.
Printer
" menu item.