Thermo Scientific Orion DUAL STAR User Manual page 93

Ph/ise
Hide thumbs Also See for Orion DUAL STAR:
Table of Contents

Advertisement

Error Code
Description
E0
Data Transfer Error – A communication error between the meter and computer
occurred. Check the cable connections and communication protocols. Make
sure that the meter is set to the correct baud rate and the computer function is
enabled in the setup menu.
E0
ncremental Techniques Error – Make sure that the correct incremental
technique was used and all of the parameters were entered correctly. The
measured millivolt difference may not be great enough to permit the calculation
of the unknown concentration. Increase the concentration of the standard or
dilute the sample. The concentration of the unknown may have been calculated
to be less than 0.0001. If the unknown was diluted, use a fresh aliquot of
undiluted or less diluted unknown. The electrode ID may be incorrect or the
electrode may not have been prepared correctly, filled with the proper filling
solution or connected to the appropriate meter input. If the error persists,
measure a standard with a known concentration and the same background as
the sample. Adjust the standard concentration and volume until accurate results
are obtained.
E0
Printer Communication Error – A communication error between the meter
and printer occurred. Check the cable connections and communication protocols.
Make sure that the meter is set to the correct baud rate for the printer and the
printer function is enabled in the setup menu. The printer may be out of paper,
the printer paper may be jammed or there may be another printer error causing
the printer not to be ready.
E0
ncorrect Password Entered – The password that was entered into the meter
was incorrect. Verify the password and reenter it.
E0
Bad Value Entered – The value that was entered into the meter was outside of
the allowable range of values. Verify the value range and reenter the value.
E0
Calibration Log Full – The calibration log is full and must be cleared before any
additional calibration data can be saved.
E0
Calibration Log Empty – There is no calibration saved in the calibration log and
a calibration needs to be performed before any calibration data can be viewed.
E00
Data Log Full – The data log is full and must be cleared before any data points
can be saved.
E0
Data Log Empty – There are no data points saved in the data log. The data log
needs to be enabled in the setup menu and a measurement needs to be logged
before any data points can be viewed.
Thermo Scientific Orion DUAL STAR
pH/ISE Meter User Guide