In the measurement mode, press the setup key . Press the ▲ / ▼ keys to
highlight Autosampler and press the f2 (select) key .
Autosampler – Press the ▲ / ▼ keys to highlight Autosampler and press
the f2 (select) key . Press the ▲ / ▼ keys to highlight On or Off to turn
communication between the meter and the autosampler on or off and press
the f2 (accept) key .
Ch .1 Method – Press the ▲ / ▼ keys to highlight Ch.1 Method and press the
f2 (select) key . To set the method that the autosampler will use for channel
1, press the ▲ / ▼ keys to highlight the method number and press the
f2 (accept) key .
Note: A method can be selected when the autosampler setup parameters
are entered, but any method editing must be done from the methods list.
Calibration values from the last calibration performed in the method are stored
to the method, so calibrations can be performed on the autosampler.
Ch .2 Method – Press the ▲ / ▼ keys to highlight Ch.2 Method and press the
f2 (select) key . To set the method that the autosampler will use for channel
2, press the ▲ / ▼ keys to highlight the method number and press the
f2 (accept) key .
Note: A method can be selected when the autosampler setup parameters
are entered, but any method editing must be done from the method list.
Calibration values from the last calibration performed in the method are stored
to the method, so calibrations can be performed on the autosampler.
Calibration – Press the ▲ / ▼ keys to highlight Calibration and press the
f2 (select) key . Press the ▲ / ▼ keys to highlight On or Off to turn the
calibration function for the autosampler on or off and press the
f2 (accept) key .
Temperature – Press the ▲ / ▼ keys to highlight Temperature and press
the f2 (select) key . Press the ▲ / ▼ keys to highlight ATC1 or ATC2 as the
temperature source of the ATC probe that will be used on the autosampler
and press the f2 (accept) key .
Tray Size – Press the ▲ / ▼ keys to highlight Tray Size and press the
f2 (select) key . Press the ▲ / ▼ keys to highlight 28 or 48 as the size of the
tray that will be used on the autosampler and press the f2 (accept) key .