Meter Setup
Read Type
Press the ▲ / ▼ keys to highlight Read Type and press the f2 (select) key .
Press the ▲ / ▼ keys to highlight Auto-Read, On Ready, At Time Intervals,
Continuous, Value Change In Measurement or Timed Reading as the read type that
will be used to display the measurements for the selected channel and press
the f2 (select) key .
Auto-Read – The meter will display the measurement as it stabilizes and
lock and hold the measurement when it is stable . Once a measurement
is locked, press the measure key to take a new measurement . Once
the measurement is locked, the meter will automatically export the
measurement to the data log, if the data log is enabled in the setup menu,
and to a printer or computer, if a printer or computer is connected to the
meter and enabled in the setup menu .
On Ready – The meter will show stabilizing on the display until the
measurement is stable . When the measurement is stable, the meter will
show ready on the display . The display will automatically update if the
measurement changes and the measurement will not lock . Once ready is
shown on the display, the meter will automatically export the measurement
to the data log, if the data log is enabled in the setup menu, and to a printer
or computer, if a printer or computer is connected to the meter and enabled
in the setup menu .
At Time Intervals – The meter will read and display a measurement at
the set time interval that is programmed by the operator . At the set time
interval, the meter will automatically export the measurement to the
data log, if the data log is enabled in the setup menu, and to a printer or
computer, if a printer or computer is connected to the meter and enabled in
the setup menu . If At Time Intervals is selected, enter the time interval value
in a hours:minutes:seconds format using the numeric keypad and press the
f2 (accept) key .