Meter Setup
v . Value Change In Measurement – The meter will display the
measurement when the reading reaches or exceeds the set high or low
value that is programmed by the operator . When the reading reaches or
exceeds the set high or low value, the meter will automatically export
the measurement to the data log, if the data log is enabled in the
setup menu, and to a printer or computer, if a printer or computer is
connected to the meter and enabled in the setup menu . If Value Change
In Measurement is selected, press the ▲ / ▼ keys to highlight Min Value,
enter the minimum measurement value using the numeric keypad and
the decimal key, press the ▲ / ▼ keys to highlight Max Value, enter
the maximum measurement value using the numeric keypad and the
decimal key and press the f3 (next) key .
vi . Timed Reading – The meter will display a measurement after a set
time delay and lock and hold the measurement after the time delay is
reached . To start another time delay cycle, press the measure key .
Once the measurement is locked, the meter will automatically export
the measurement to the data log, if the data log is enabled in the
setup menu, and to a printer or computer, if a printer or computer is
connected to the meter and enabled in the setup menu . If Timed Reading
is selected, enter the time delay value in a minutes:seconds format
using the numeric keypad and press the f2 (accept) key .
4 . Repeat step 3 for channel 2 .
5 . Press the ▲ / ▼ keys to highlight Printer, Computer or None as where the meter
will export the measurement and calibration data and press the f2 (accept)
key . If Printer or Computer is selected, the meter will prompt the operator to set
the baud rate . Press the ▲ / ▼ keys to highlight 1200, 2400, 4800, 9600, 19200 or
38400 as the baud rate and press the f2 (accept) key .
6 . Press the f2 (yes) key to enable the exporting of the measurement data
to the data log or press the f3 (no) key to disable the exporting of the
measurement data to the data log .
7 . Press the f2 (measure) key to save the EZ Startup menu settings and
proceed to the measurement mode or press the f1 (back) key to edit any of
the EZ Startup settings .